@Adam Equipment Company 2005
7
4.1 CALIBRATION
•
Calibration Mass required:
ADK-10
: 10kg
ADK-20
: 20kg
•
Press
the
[On/Off]
switch in the rear of the balance to turn the
balance on.
•
Display will show a self-check mode.
•
After about 5 seconds, zero will be displayed.
•
Warm up at least one hour before calibrating.
•
Make sure that nothing is on the pan.
•
Press
[Tare]
key to zero the display.
•
Display will show “
0.0
” - If the display is not zeroed, press the
[Tare]
key again.
•
Press
the
[Cal]
key.
•
Display will show “
CAL 0.0
”
•
If necessary press
[Tare]
again and then press
[Cal]
.
•
Place the calibration mass on to the pan gently (Take ADK-20 for
example).
•
Press
[Cal]
again.
•
The display will show “
C20000.0
” and then “
CAL End
” will be displayed.
•
Remove the mass from the pan gently.
•
Display will show “
0.0
”. The balance now comes into weighing mode.
•
Note: if “
no CAL
” is displayed during calibration, it means that the
balance can't be calibrated.