40
|
P a g e
© Adam Equipment Company 2011
12.3.4 F4 –INITIAL ZERO RANGE
To enter this parameter, press the [Z/T] key when “
F4 int
” is shown.
The display will show the current initial zero range.
Press the [Pcs/ ] key to change the value and press [Z/T] to accept the
value.
Press [Print/M+/esc] to return to weighing.
12.3.5 F5 -RE-ZERO RANGE
To enter this parameter, press the [Z/T] key when “
F5 rEZ
” is shown.
The display will show the current re-zero range.
Press the [Pcs/ ] key to change the value.
Press [Z/T] to accept the value.
Press [Print/M+/esc] to return to weighing.
12.3.6 F6 -SUCCESSIVE TARE
To enter this parameter, press the [Z/T] key when “
F6 SCS
” is shown.
The display will show if the successive tare is on or off.
Press the [Pcs/ ] key to change the value.
Press [Z/T] to accept the displayed value.
Press [Print/M+/Esc] to return to weighing.
12.3.7 F7 –ADC COUNTS
To enter this parameter, press the [Z/T] key when “
F7 Cnt
” is shown.
This parameter allows you to view the A/D counts from the internal A/D
converter. This can be an aid to service.
Press the [Z/T] key to return to the PARAMETER menu.
Press the [Print/M+/Esc] key to return to weighing.
Typical value at zero is 30,000-90,000 (approx.).
Typical value at full capacity is 500,000 (approx.).
12.3.8 F8 –ZERO MODE
To enter this parameter, press the [Z/T] key when “
F8 ZEm
” is shown.
Select the Zero mode desired. In all but special cases Zero Mode 1 is used.
The other 2 zero modes are for unique locations in the world and effect the
+/- range of the zero.
Press the [Pcs/ ] key to change the value.
Press [Z/T] to accept the displayed value.
Press [Print/M+/Esc] to return to weighing.