25
Device Handling
Switching On and Off
Switching the device on and off
The on/off key is located on the right side at the rear of the device.
WARNING
Risk of energy being conducted to the patient
If the device and the patient are touched at the same time, electrical
energy can be conducted from the device into the patient.
• Never touch the device and the patient at the same time.
Note: While the AcQBlate® FORCE ablation catheter and a RF generator
are connected to Qubic Force it is always possible to start a cardiac radio
frequency ablation, even if there is an error in how the contact force is
displayed or if the Qubic Force is switched off.
• To switch the device on or off, press the on/off key.
After switching on the device, the on/off light indicator on the front
left lights up and Qubic Force performs a self-test. After the self-test,
the main view appears on the external monitor.
• To disconnect Qubic Force from the mains supply, pull the power plug
of the device.