Device Handling
• Connect the PK-147 cable to the red Redel port on the device.
• Connect the PK-147 cable to the AcQBlate® FORCE ablation catheter.
• Connect the optical plug on the ablation catheter to the optical port
on the device.
Refer to the technical manual of the ablation catheter.
Once connected, it may take up to 10 s before the ablation catheter can be
used.
The first connection of the AcQBlate® FORCE ablation catheter to Qubic
Force is stored and, from this time, the AcQBlate® FORCE ablation catheter
can be used for 24 hours. During this time, you can remove the AcQBlate®
FORCE ablation catheter from the device, for example.
The values used to obtain contact force and application angle are
automatically stored by the device upon first connecting the AcQBlate®
FORCE ablation catheter and each time the Qubic Force is started.
Connecting the RF generator
CAUTION
There is a risk of exceeding the leakage current limits when
connecting external devices that have their own power supply
as well as a risk of making an electrically conductive connec-
tion to other devices.
• Connect to the covered blue Redel port for the RF generator only
devices that comply with IEC 60601-2-2 standard and are CF-type
applied parts.
• Before initial commissioning, check and document all device combi-
nations according to IEC standard 60601-1.
• Perform this inspection at least once per year according to the legal
requirements.