CLEANING/DISINFECTING AND FILTER REPLACEMENT INSTRUCTIONS

The ActivePure® Medical Guardian, model F170A is a device intended for medical purposes that is used for the reduction of staphylococcus 
epidermidis and erwinia herbicola bacteria, MS2 and Phi-X174 viruses and aspergillum niger fungal spores and bacillus globigii bacterial spores from 
the air in a temperature-controlled professional healthcare environment of 70~71°F, 40~45%RH.

The following instructions cover device cleaning, disinfecting and replacement of the filter. Thoroughly read all of the instructions before starting. It 
is recommended to clean and disinfect the device each time the filter is replaced or when device becomes visibly soiled.

  

 

FILTER REPLACEMENT

Metrex CaviWipes are required to clean the device when replacing the filter, but they are not provided with the unit or filter.

Step 1:

  

Always

 disconnect (unplug) device from electrical power source before cleaning, disinfecting or replacing the filter.

Step 2:

  If necessary, relocate the device to an appropriate area to perform the cleaning, disinfecting and filter replacement. It is not recommended 

to perform these tasks in areas such as patient room, operating room, examination room, etc., where patient exposure may occur. 

Step 3:

  Open the replacement filter carton which includes the following items inside. 

•  1 replacement filter in a sealed wrapper

•  1 disposable surgical mask

•  1 pair disposable latex free gloves, large

•  1 disposable 8-quadrant wiping cloth

•  1 large plastic bag

•  1 red bio hazard sticker

•  1 plastic bag seal

•  1 printed copy of instruction for use, cleaning, disinfecting, filter replacement sheet 

Step 4:

  Before removing and handling the used filter, put on the disposable mask and gloves. 

Step 5:

  Slide (remove) the used filter from the device, placing it inside the large plastic bag provided for disposal, 

then set it aside. 

Step 6:

  Use Metrex CaviWipes to clean the device. These are 

NOT

 provided with the device or replacement filter.

NOTE:

   Metrex CaviWipes are required to clean the device when replacing the filter, but they are not provided with the 

unit or filter.

Step 7:

  Remove a Metrex CaviWipe from its package. Beginning by cleaning the surface “A”. Wipe the CaviWipe over the surface area, 

thoroughly wetting the soil. Then using hand pressure rub the wipe over the surface until soil is no longer visible on this part of 
the device. If soil is heavy or is still visible, repeat step with a new CaviWipe until soil is no longer visible. 

Step 8:

  Remove a Metrex CaviWipe from its package. Beginning by cleaning the surface “B”. Wipe the CaviWipe over the surface area, 

thoroughly wetting the soil. Then using hand pressure rub the wipe over the surface until soil is no longer visible on this part of 
the device. If soil is heavy or is still visible, repeat step with a new CaviWipe until soil is no longer visible.

Step 9:

  Remove a Metrex CaviWipe from its package. Beginning by cleaning the surface “C”. Wipe the CaviWipe over the surface area, 

thoroughly wetting the soil. Then using hand pressure rub the wipe over the surface until soil is no longer visible on this part of 
the device. If soil is heavy or is still visible, repeat step with a new CaviWipe until soil is no longer visible.

1

2

3

4

5

A

B

C

Summary of Contents for F170A

Page 1: ...1 READ MANUAL CAREFULLY FOR PROPER USE AND OPTIMAL OPERATION MODEL F170A The Aberdeen Bldg 14841 Dallas Parkway Suite 500 Dallas TX 75254...

Page 2: ...eep cord away from heated surfaces Do not unplug by pulling on cord to unplug grasp the plug to remove not the cord Never handle plug or device with wet hands Keep hair loose clothing fingers and all...

Page 3: ...vice Specifications 6 Operation 6 Treatment Space 7 Settings 8 General Maintenance Instructions 8 Cleaning Disinfecting and Filter Replacement Instructions 9 ActivePure Cell Replacement Instructions 1...

Page 4: ...cuit board which controls the device operations and an internal fan to draw air through the device as shown in the general device design Figure 1 Lab test results demonstrated the ActivePure Medical G...

Page 5: ...atively charged particles which flow out from a series of carbon brushes Carbon to remove odors Hydrophobic polypropylene pleated media filter designed to reduce airborne contaminants Powerful UVGI bu...

Page 6: ...h 300 CFM 509 m3 h OPERATION DEVICE SPECIFICATIONS Place the ActivePure Medical Guardian in a location that allows air to move freely into around and out of the unit Allow minimum of 12 of clearance a...

Page 7: ...e Movement of people or equipment within the space Rates of ventilation The device is intended to result in a 4 6 log reduction of airborne microorganisms from initial concentrations in 60 minutes as...

Page 8: ...r of the cabinet with a clean soft cloth as needed to remove finger prints smudges etc Clean and disinfect device after each filter change as detailed in the Cleaning Disinfecting and Filter Replaceme...

Page 9: ...hazard sticker 1 plastic bag seal 1 printed copy of instruction for use cleaning disinfecting filter replacement sheet Step 4 Before removing and handling the used filter put on the disposable mask an...

Page 10: ...of the device If soil is heavy or is still visible repeat step with a new CaviWipe until soil is no longer visible Step 15 Allow the device to dry no visible wetting remains No rinsing is required Wi...

Page 11: ...at is used for the reduction of staphylococcus epidermidis and erwinia herbicola bacteria MS2 and Phi X174 viruses and aspergillum niger fungal spores and bacillus globigii bacterial spores from the a...

Page 12: ...des longer fan life Q Why doesn t the display dim immediately when the room gets dark A The automatic dim feature has a brief delay to prevent unnecessary dimming if a shadow momentarily passes in fro...

Page 13: ...use ActivePure Medical s trademarked names images and logos as well as Internet auction sites e g ebay and Craigslist The only approved Internet presence for ActivePure Medical LLC products is www act...

Page 14: ...61000 4 11 This device conforms fully to the EMC requirement of IEC 60601 1 2 and is suitable for use in a professional healthcare environment This device is not effected by RFID wireless networks 2...

Page 15: ...15...

Page 16: ...vePure Medical LLC The Aberdeen Bldg 14841 Dallas Parkway Suite 500 Dallas TX 75254 1 800 572 0317 ActivePure Medical LLC Canada Inc 3480 Laird Road Suite 2 5 Mississauga ON L5L 5Y4 For information re...

Reviews: