DEVICE IEC 60601-1-2:2014 EMC INFORMATION

This device passed each of the 11 tests, as indicated. Immunity test levels performed on the device,

•  Electrostatic Discharge Immunity –

 Tested at 2,4,8 and 15 kV

•  Radiated Immunity – 

Tested 3 V/m (Professional Healthcare) and 10 V/m (Home Healthcare)

•  Conducted Immunity –

 Tested 150 Hz ~ 80 MHz with AM 80% @ 1kHz 3V and 6V

TEST STANDARD

Conducted Emissions (EN 55011, CISPR 11)

Radiated Emissions (EN 55011, CISPR 11)

Harmonics (EN 61000-3-2)

Flicker (EN 61000-3-3: IEC 61000-4-2)

Electrostatic Discharge (EN 61000-4-2: IEC 6100-4-2)

Radiated Immunity (EN 61000-4-3: IEC 61000-4-3)

Fast Transient Burst (EN 61000-4-4: IEC 61000-4-4)

Surges (EN 61000-4-5: IEC 61000-4-5)

Conducted Immunity (EN 61000-4-6: IEC 61000-4-6)

Power Frequency Magnetic Field (EN 61000-4-8: IEC 61000-4-8)

Voltage Dips and Interrupt (EN 61000-4-11: IEC 61000-4-11)

•  This device conforms fully to the EMC requirement of IEC 60601-1-2 and is suitable for use in a professional healthcare environment

•  This device is not effected by RFID, wireless networks, 2-ways radios, paging systems, etc., as the device does not transmit or receive RF 

signals or create EMF beyond the required limits

•  Do not use device with an extension cord as unintended EMF may be created

•  This device is grounded to prevent electrostatic discharge

•  Do not use RF communication or magnetic field generating equipment within 30 cm of the device

•  EMC testing shows no conducted or radiated electromagnetic emissions or immunity issues which would be adverse to the patient or 

operator

The use of this device in a manner other than described in the user manual or a modification of the device could result in increased 
electromagnetic emissions or decreased immunity.

If device is operating in an unexpected manner or causing unexpected interference , discontinue using the device, review the user 
manual or seek service.

Summary of Contents for F170A

Page 1: ...1 READ MANUAL CAREFULLY FOR PROPER USE AND OPTIMAL OPERATION MODEL F170A The Aberdeen Bldg 14841 Dallas Parkway Suite 500 Dallas TX 75254...

Page 2: ...eep cord away from heated surfaces Do not unplug by pulling on cord to unplug grasp the plug to remove not the cord Never handle plug or device with wet hands Keep hair loose clothing fingers and all...

Page 3: ...vice Specifications 6 Operation 6 Treatment Space 7 Settings 8 General Maintenance Instructions 8 Cleaning Disinfecting and Filter Replacement Instructions 9 ActivePure Cell Replacement Instructions 1...

Page 4: ...cuit board which controls the device operations and an internal fan to draw air through the device as shown in the general device design Figure 1 Lab test results demonstrated the ActivePure Medical G...

Page 5: ...atively charged particles which flow out from a series of carbon brushes Carbon to remove odors Hydrophobic polypropylene pleated media filter designed to reduce airborne contaminants Powerful UVGI bu...

Page 6: ...h 300 CFM 509 m3 h OPERATION DEVICE SPECIFICATIONS Place the ActivePure Medical Guardian in a location that allows air to move freely into around and out of the unit Allow minimum of 12 of clearance a...

Page 7: ...e Movement of people or equipment within the space Rates of ventilation The device is intended to result in a 4 6 log reduction of airborne microorganisms from initial concentrations in 60 minutes as...

Page 8: ...r of the cabinet with a clean soft cloth as needed to remove finger prints smudges etc Clean and disinfect device after each filter change as detailed in the Cleaning Disinfecting and Filter Replaceme...

Page 9: ...hazard sticker 1 plastic bag seal 1 printed copy of instruction for use cleaning disinfecting filter replacement sheet Step 4 Before removing and handling the used filter put on the disposable mask an...

Page 10: ...of the device If soil is heavy or is still visible repeat step with a new CaviWipe until soil is no longer visible Step 15 Allow the device to dry no visible wetting remains No rinsing is required Wi...

Page 11: ...at is used for the reduction of staphylococcus epidermidis and erwinia herbicola bacteria MS2 and Phi X174 viruses and aspergillum niger fungal spores and bacillus globigii bacterial spores from the a...

Page 12: ...des longer fan life Q Why doesn t the display dim immediately when the room gets dark A The automatic dim feature has a brief delay to prevent unnecessary dimming if a shadow momentarily passes in fro...

Page 13: ...use ActivePure Medical s trademarked names images and logos as well as Internet auction sites e g ebay and Craigslist The only approved Internet presence for ActivePure Medical LLC products is www act...

Page 14: ...61000 4 11 This device conforms fully to the EMC requirement of IEC 60601 1 2 and is suitable for use in a professional healthcare environment This device is not effected by RFID wireless networks 2...

Page 15: ...15...

Page 16: ...vePure Medical LLC The Aberdeen Bldg 14841 Dallas Parkway Suite 500 Dallas TX 75254 1 800 572 0317 ActivePure Medical LLC Canada Inc 3480 Laird Road Suite 2 5 Mississauga ON L5L 5Y4 For information re...

Reviews: