ActivePure F170A Manual Download Page 11

11

CLEANING/DISINFECTING AND FILTER REPLACEMENT INSTRUCTIONS 

(CONT.)

Step 23:

  Spray disinfecting solution over surface G. Wait 4 minutes, ± 5 seconds, then thoroughly wipe with the cloth folded to quadrant 7. Continue wiping 

until all solution has been removed. 

Step 24:

  Spray disinfecting solution over surface H. Wait 4 minutes, ± 5 seconds, then thoroughly wipe with the cloth folded to quadrant 8. Continue wiping 

until all solution has been removed.

Step 25:

  When finished, place the disposable mask, gloves and wiping cloth in the large plastic bag with the used filter. 

Step 26:

  Secure the large plastic bag with the provided bag closure and place the red biohazard label on the plastic bag. Disposal to be in accordance with 

facility policy.  

Step 27:

  Remove the new filter from its protective wrap and slide (following directions on filter) it into the device. The protective filter wrap is not a biohazard, 

so no special handling is required.

Step 28:

  Allow the device to dry, no visible wetting remains, and no rinsing is required. Plug device in and perform a reset of the filter change light by pressing 

and holding the power button for 10 seconds until the filter replace light  goes out.

Step 29:

  The device is now ready to be returned to use.

ACTIVEPURE® CELL REPLACEMENT INSTRUCTIONS

The ActivePure® Medical Guardian, model F170A is a device intended for medical purposes that is used for the reduction of staphylococcus epidermidis and 
erwinia herbicola bacteria, MS2 and Phi-X174 viruses and aspergillum niger fungal spores and bacillus globigii bacterial spores from the air, in a temperature-
controlled professional healthcare environment of 70~71°F, 40~45% RH. 

The following instructions will permit the device ActivePure Cell to be safely replaced without creating safety concerns in the patient 
environment. 

Step 1:

  Always disconnect (unplug) device from electrical power source before replacing the ActivePure Cell.

Step 2:

  If necessary, relocate the device to an appropriate area to perform cell replacement. Do not replace cell in areas such as patient room, operating room, 

examination room, etc., where patient exposure may occur. 

Step 3:

  Remove the 4 screws securing the old cell mounting plate on the device.

Step 4:

  Slide the old cell out of the device and disconnect the two bulb sockets.

Step 5: 

 Open the replacement cell carton and remove the new cell.

Step 6:

  Place old cell inside the empty new cell carton and dispose of properly.

NOTE:

   No special handling for contaminants is necessary as the cell is on the exhaust side of the device and does not trap or contain contaminants.

Step 7:

  Reattach the two bulb sockets to the new cell.

Step 8:

  Slide the new cell back inside the device and secure with the 4 screws removed in Step 3.

Step 9:

  Plug device in and reset the ActivePure Cell Change Light by pressing and holding the power button for 10 seconds until it goes out.

Step 10: 

 The device is now ready to be returned to its location.

Summary of Contents for F170A

Page 1: ...1 READ MANUAL CAREFULLY FOR PROPER USE AND OPTIMAL OPERATION MODEL F170A The Aberdeen Bldg 14841 Dallas Parkway Suite 500 Dallas TX 75254...

Page 2: ...eep cord away from heated surfaces Do not unplug by pulling on cord to unplug grasp the plug to remove not the cord Never handle plug or device with wet hands Keep hair loose clothing fingers and all...

Page 3: ...vice Specifications 6 Operation 6 Treatment Space 7 Settings 8 General Maintenance Instructions 8 Cleaning Disinfecting and Filter Replacement Instructions 9 ActivePure Cell Replacement Instructions 1...

Page 4: ...cuit board which controls the device operations and an internal fan to draw air through the device as shown in the general device design Figure 1 Lab test results demonstrated the ActivePure Medical G...

Page 5: ...atively charged particles which flow out from a series of carbon brushes Carbon to remove odors Hydrophobic polypropylene pleated media filter designed to reduce airborne contaminants Powerful UVGI bu...

Page 6: ...h 300 CFM 509 m3 h OPERATION DEVICE SPECIFICATIONS Place the ActivePure Medical Guardian in a location that allows air to move freely into around and out of the unit Allow minimum of 12 of clearance a...

Page 7: ...e Movement of people or equipment within the space Rates of ventilation The device is intended to result in a 4 6 log reduction of airborne microorganisms from initial concentrations in 60 minutes as...

Page 8: ...r of the cabinet with a clean soft cloth as needed to remove finger prints smudges etc Clean and disinfect device after each filter change as detailed in the Cleaning Disinfecting and Filter Replaceme...

Page 9: ...hazard sticker 1 plastic bag seal 1 printed copy of instruction for use cleaning disinfecting filter replacement sheet Step 4 Before removing and handling the used filter put on the disposable mask an...

Page 10: ...of the device If soil is heavy or is still visible repeat step with a new CaviWipe until soil is no longer visible Step 15 Allow the device to dry no visible wetting remains No rinsing is required Wi...

Page 11: ...at is used for the reduction of staphylococcus epidermidis and erwinia herbicola bacteria MS2 and Phi X174 viruses and aspergillum niger fungal spores and bacillus globigii bacterial spores from the a...

Page 12: ...des longer fan life Q Why doesn t the display dim immediately when the room gets dark A The automatic dim feature has a brief delay to prevent unnecessary dimming if a shadow momentarily passes in fro...

Page 13: ...use ActivePure Medical s trademarked names images and logos as well as Internet auction sites e g ebay and Craigslist The only approved Internet presence for ActivePure Medical LLC products is www act...

Page 14: ...61000 4 11 This device conforms fully to the EMC requirement of IEC 60601 1 2 and is suitable for use in a professional healthcare environment This device is not effected by RFID wireless networks 2...

Page 15: ...15...

Page 16: ...vePure Medical LLC The Aberdeen Bldg 14841 Dallas Parkway Suite 500 Dallas TX 75254 1 800 572 0317 ActivePure Medical LLC Canada Inc 3480 Laird Road Suite 2 5 Mississauga ON L5L 5Y4 For information re...

Reviews: