INSTRUCTION MANUAL
6
EN
TROUBLESHOOTING
The ALERT J4X breath alcohol tester requires calibration each year to maintain
accuracy. When calibration is due, CAL is displayed.
INCIDENT / MESSAGE
RESOLUTION
No Power
Remove the plastic tab from the battery compartment
No Test
Review the “Taking a Breath Test” section. Contact ACS
for assistance
CAL
Calibration is due. Contact ACS for service
E01
Change the batteries
E02-E05, E15, E16
Power the tester off and power back on. If the problem
persists, contact ACS
E06 - E08
Temperature error - the ALERT J4X is outside the
temperature range (-5 to 40 °C) and cannot operate in
accordance with the standard EN 15 964
E09-E25
A self-diagnosis procedure has detected an error in the
internal operation. Contact ACS for service
WARRANTY
The ALERT J4X breath alcohol tester is warranted against defects in materials or
workmanship for 1 year from the date of purchase. During the warranty period,
Alcohol Countermeasure Systems agrees to replace or repair any unit or part that in
its judgement is found to be defective, providing the defect is not due to damage or
misuse.
SERVICE
Units requiring service should be packed securely and shipped prepaid to an
authorized Service Centre.
ORDERING MOUTHPIECES
Contact ACS or visit shopacs.com for Mouthpieces (Bag of 25), ACS Part Number:
95-000145 or 95-000141.
Summary of Contents for ALERT J4X
Page 1: ...Instruction manual B R E AT H A L C O H O L T E S T E R...
Page 4: ......
Page 5: ...English 1 Italiano 10 Nederlands 20 Portugu s 30 Espa ol 40 Svensk 50...
Page 6: ......
Page 15: ...EN...
Page 16: ......
Page 26: ......
Page 35: ...NL...
Page 36: ......
Page 45: ...PT...
Page 46: ......
Page 55: ......
Page 56: ......
Page 65: ......
Page 66: ......
Page 67: ......
Page 68: ...IRM MULTI 60 000743 A 2015...