OWL Data Loggers
27
Copyright © 2010 ACR Systems Inc. All Rights Reserved.
3.
Select a range that best fits the expected data - see Table 4-1. DC Voltage ranges are
picked to match the transducer's full scale output value. For example, if a transducer with
a 0 to 5 volt output is used then range number 18 which measures up to 7.3 VDC gives
the best resolution.
4.
Refer to the
Figure 4.1: DC Voltage Wiring
Table 4.1: DC Voltage Ranges
Range
Full Scale
Range
Full Scale
0
121.0 mVDC
16
2.431 VDC
1
241.9 mVDC
17
4.863 VDC
2
362.9 mVDC
18
7.294 VDC
3
483.9 mVDC
19
9.73 VDC
4
605.8 mVDC
20
12.18 VDC
5
726.8 mVDC
21
14.61 VDC
6
848 mVDC
22
17.04 VDC
7
969 mVDC
23
19.47 VDC
8
1.093 VDC
24
21.96 VDC
9
1.214 VDC
25
24.39 VDC
10
1.334 VDC
26
26.82 VDC
11
1.455 VDC
27
29.25 VDC
12
1.577 VDC
28
31.71 VDC
13
1.698 VDC
29
34.14 VDC
14
1.819 VDC
30
36.57 VDC
Summary of Contents for OWL 100
Page 3: ...TrendReader is a Registered Trademark of ACR Systems Inc TRADEMARK NOTICES ...
Page 4: ......
Page 44: ...OWL Data Loggers 37 Copyright 2010 ACR Systems Inc All Rights Reserved ...
Page 45: ...OWL Data Logger 38 Copyright 2010 ACR Systems Inc All Rights Reserved ...
Page 46: ...OWL Data Loggers 39 Copyright 2010 ACR Systems Inc All Rights Reserved ...
Page 48: ...OWL Data Loggers 41 Copyright 2010 ACR Systems Inc All Rights Reserved ...
Page 49: ...OWL Data Logger 42 Copyright 2010 ACR Systems Inc All Rights Reserved ...
Page 50: ...OWL Data Loggers 43 Copyright 2010 ACR Systems Inc All Rights Reserved ...
Page 54: ......
Page 55: ......