OWL Data Logger
12
Copyright © 2010 ACR Systems Inc. All Rights Reserved.
1.2.2.2
Logger Memory
The Memory subsection indicates how the data is stored in memory and what the start and
end times (or capacity) are for the data logging. In the FIFO (First In, First Out) mode, the
logger continually stores data and when the memory limit is reached, the new readings start to
overwrite the oldest readings. Using this method, the logger is always collecting data, but only
saves the most recent data in its memory. In the Stop When Full mode, when the memory
limit is reached, the logger stops recording.
The length of time for data capture of the logger is determined by the sampling rate. To find
out more about how the capacity is affected by the sample rate, see
.
Usage
To set how the logger stores its data, click on the Usage field and select FIFO or Stop When
Full.
Start Time
To set when the logger will start recording click on the Start Time field. Next, click on the
small button that appears at the right. This will pop up a form allowing you to set the date and
time. Highlight the date or time and enter the appropriate value. You can also click on the
drop-down list arrow to bring up a calendar. You can set the logger's starting time for up to 6
Summary of Contents for OWL 100
Page 3: ...TrendReader is a Registered Trademark of ACR Systems Inc TRADEMARK NOTICES ...
Page 4: ......
Page 44: ...OWL Data Loggers 37 Copyright 2010 ACR Systems Inc All Rights Reserved ...
Page 45: ...OWL Data Logger 38 Copyright 2010 ACR Systems Inc All Rights Reserved ...
Page 46: ...OWL Data Loggers 39 Copyright 2010 ACR Systems Inc All Rights Reserved ...
Page 48: ...OWL Data Loggers 41 Copyright 2010 ACR Systems Inc All Rights Reserved ...
Page 49: ...OWL Data Logger 42 Copyright 2010 ACR Systems Inc All Rights Reserved ...
Page 50: ...OWL Data Loggers 43 Copyright 2010 ACR Systems Inc All Rights Reserved ...
Page 54: ......
Page 55: ......