- 14 -
OMNISTIM
®
FX
2
PORTABLE PENS PATIENT USER MANUAL
COPYRIGHT © 2008 - 2021, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED
CLEANING
Cleaning / Disinfecting of the OMNISTIM
®
FX
2
PORTABLE PENS
It is recommended to clean the device and lead wire attachments after each use to control the
spread of infection.
•
Clean the Omnistim
®
FX
2
Portable PENS after each use with ACP germicidal wipes. Wipe
common contact surfaces, such as control panel, lead wires, and probe tips with germicidal
disposable wipes and allow to air dry. This technique will inactivate most bacteria and viruses.
This will also facilitate removal of contaminants from the equipment and accessories.
•
Disposable/reusable electrodes are for individual patient use only and should not be shared
with others.
•
All disposable electrodes should be discarded after each use. Do not attempt to clean and
reuse disposable electrodes.
TROUBLESHOOTING
The following table lists machine problem symptoms and possible areas to check for the problem
causes. If these suggested measures do not correct the machine malfunction, call your physician
for assistance.
PROBLEM
CAUSE
REMEDY
Unit will not power on
•
No batteries
•
Low batteries
•
Install batteries
•
Replace batteries
•
Verify type of batteries
•
Inspect battery contacts
Display shows low
battery symbol
•
Battery voltage is too low
•
Replace batteries for future use
Channel indicator
flashing and caution
symbol is displayed
•
Batteries are too low to perform treatment
•
Lead wires not properly connected
•
Lead wire to electrode not connected
•
Improper skin preparation
•
Replace batteries
•
Replace lead wire(s) if defective
•
Properly connect leads as needed
•
Properly prepare skin prior to treatment
Patient feels surging or
spiking sensation
•
Lead wire(s) breakage
•
Electrode(s) lead wire breakage
•
Replace lead wire(s) if defective
•
Replace electrode(s)
Patient cannot detect
output
•
Failure of lead wire(s)
•
Electrode(s) failure, or electrode gel
dried out
•
Device failure
•
Replace lead wire(s) if defective
•
Remove electrode(s) and replace
if necessary
•
Contact ACP customer service
Symbols
Caution or Attention. There is a problem. No patient detected.
Output locked. Decrease output to unlock then increase output.
Reduce output before increasing the channel to desired level.
Summary of Contents for OMNISTIM FX2
Page 2: ......