- 24 -
OMNISTIM
®
500 PRO USER MANUAL
COPYRIGHT © 2004 - 2018, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED
B. MFAC TONE-SPASM:
Relaxation of muscle spasm. Target is superficial and deep muscle groups. Bipolar set-up over agonist and
antagonist muscles for target muscle. Intensity should be set to elicit a grade 2-3 muscle contraction.
Activation of muscle with electrical stimulation with a 1:1 ratio of “ON/OFF” times causes muscle fatigue
reducing muscle spasm. Sensory stimulation of the muscles also reduces pain and thus reflex muscle
spasm.
USER ADJUSTABLE
PARAMETER (Y/N)
ORDER IN
PARAMETER
SET LOOP
PARAMETER NAME
ADJUSTABLE
RANGE (UNIT)
DEFAULT (UNIT)
Y
9
CARRIER FREQ
2.0, 2.5, 4.0, 5.0, 10.0 (kHz)
5.0 (kHz)
Y
6
BURST FREQUENCY
0.0…250 (Hz)
90 (Hz)
N
BURST RATE SCAN
20 (%)
N
RATE SCAN
ON
N
RATE SCAN TIME
4 (s)
Y
1
TREATMENT TIME
0…99 (min)
15 (min)
Y
2
8
2
B OUTPUT DELAY
0.0…9.9 (s)
0.0 (s)
Y
4
RAMP UP
0.0…9.9 (s)
2.0 (s)
Y
5
RAMP DOWN
0.0…9.9 (s)
2.0 (s)
Y
2
T-ON
0…30 (s)
10 (s)
1
Y
3
T-OFF
0…199 (s)
10 (s)
1
Y
7
(93h) Channel Mode
A/B SIMULTANEOUS, A/B
ALTERNATE, B DELAYED
A/B SIM.
1
When T-OFF = 0, T-ON = CONTINUOUS.
2
B OUTPUT DELAY parameter is shown only when the Channel Mode is set to B DELAYED.
C. MFAC RECIPROCAL INHIBITION:
Reduces muscle tone (spasm) of spastic muscle. A single channel (A or B) is used over the spastic
muscle’s antagonist. Set intensity to elicit a grade 2-3 muscle contraction. Ensure the intensity is not too
high to induce overflow activation of the spastic muscle. Reduce treatment time based on muscle fatigue
and inhibition of spasticity. Stimulation of the spastic muscle’s antagonists activates reciprocal inhibition
of the spastic muscle reducing tone. Slow ramps decrease the potential to trigger spasticity of the agonist.
USER ADJUSTABLE
PARAMETER (Y/N)
ORDER IN
PARAMETER
SET LOOP
PARAMETER NAME
ADJUSTABLE
RANGE (UNIT)
DEFAULT (UNIT)
Y
2
CARRIER FREQ
2.0, 2.5, 4.0, 5.0, 10.0 (kHz)
2.5 (kHz)
Y
3
BURST FREQUENCY
0.0…250 (Hz)
50 (Hz)
N
BURST RATE SCAN
20 (%)
N
RATE SCAN
ON
N
RATE SCAN TIME
4 (s)
Y
1
TREATMENT TIME
0…99 (min)
15 (min)
Y
2
9
2
B OUTPUT DELAY
0.0…9.9 (s)
0.0 (s)
Y
6
RAMP UP
0.0…9.9 (s)
6.0 (s)
Y
7
RAMP DOWN
0.0…9.9 (s)
4.0 (s)
Y
4
T-ON
0…30 (s)
12 (s)
1
Y
5
T-OFF
0…199 (s)
18 (s)
1
Y
8
Channel Mode
A/B SIMULTANEOUS, A/B
A/B SIM.
1
When T-OFF = 0, T-ON = CONTINUOUS.
2
B OUTPUT DELAY parameter is shown only when the Channel Mode is set to B DELAYED.