- 18 -
OMNISTIM
®
500 PRO USER MANUAL
COPYRIGHT © 2004 - 2018, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED
OMNISTIM
®
500 PRO PROGRAM MENU
Pain Control
The following protocols are found by pressing the PAIN protocol button. The first protocol displayed is IFC
SENSORY. Other protocols follow, in the sequence presented below, by continuously pressing the PAIN button.
A. IFC SENSORY:
Provides symptomatic relief of superficial and deep pain from localized dermatome or segmental origin.
Bilateral, bipolar or quadripolar application through the painful area or over the involved spinal segments.
Apply parallel to incision line for post op pain management. Set intensity to illicit a pleasant tingling
sensation, just below muscle contraction. Sensory stimulation activates A-beta fibers causing the release of
spinal Enkephalin and Dynorphin to block pain at the segmental level (Gate Control). Duration of relief is
typically from 30 minutes to 2 hours. Fast onset of relief is usually within 15-20 minutes.
USER ADJUSTABLE
PARAMETER (Y/N)
ORDER IN
PARAMETER
SET LOOP
PARAMETER NAME
ADJUSTABLE
RANGE (UNIT)
DEFAULT
(UNIT)
N
CARRIER FREQ
5.0 (kHz)
N
BURST FREQUENCY
100 (Hz)
N
BURST RATE SCAN
20 (%)
N
RATE SCAN
ON
N
RATE SCAN TIME
20 (s)
Y
2
VECTOR
Slow 45, Fast 45,
Slow 90, Fast 90, OFF
Fast 90
Y
1
TREATMENT TIME
0…99 (min)
15 (min)
B. TENS-LVPC SENSORY:
This program provides symptomatic relief of superficial pain from localized dermatome or segmental
origin. Bilateral, bipolar application through the painful area or over the involved spinal segments. Apply
parallel to incision line for post op pain management. Set intensity to illicit a pleasant tingling sensation,
just below muscle contraction. Sensory stimulation activates A-beta fibers causing the release of spinal
Enkephalin and Dynorphin to block pain at the segmental level (Gate Control). Duration of relief is
typically from 30 minutes to 2 hours. Fast onset of relief is usually within 15-20 minutes.
USER ADJUSTABLE
PARAMETER (Y/N)
ORDER IN
PARAMETER
SET LOOP
PARAMETER NAME
ADJUSTABLE
RANGE (UNIT)
DEFAULT
(UNIT)
Y
2
PHASE DURATION
40…300 (μs)
80 (μs)
Y
3
PULSE RATE
0.0…250 (Hz)
100 (Hz)
N
PHASE DURATION MOD
20 (%)
N
MODULATION
OFF
N
RATE SCAN TIME
4 (s)
Y
1
TREATMENT TIME
0…99 (min)
15 (min)
N
CHANNEL MODE
A/B SIM.