Specifications
Feature
Result Calibration
Sample
Minimum Sample Size
Test Time
On/Off Source
Battery Life
Glucose Units of Measure
Memory
Automatic shutoff
Meter Size
Display Size
Weight
Operating Temperature
Operating Relative Humidity
Hematocrit Range
Data Port
Measurement Range
Specification
Plasma-equivalent
Fresh whole blood (Capillary, Venous,
Neonatal, and Arterial)
0.8
µ
L
5 seconds
Two (2) CR 2032 3.0V coin cell batteries
Minimum of 3,000 measurements
(without considering data transfer
and test reminder alarm)
The meter is pre-set to either millimoles
per liter (mmol/L) or milligrams per
deciliter (mg/dL) depending on the
standard of your country
Up to 300 records with time and date
2 minutes after last action
90mm x 58mm x 21mm
41mm x 37mm
66g (with battery installed)
o
5 - 45 C
10 - 90%
20 - 70%
9600 baud, 8 data bits, 1 stop bit,
no parity
0.6-33.3 mmol/L (10 - 600 mg/dL)
En
En
Warranty
Please complete the warranty card that came with this product and mail
it to your dealer to register your purchase.
If the meter fails to work for any reason other than obvious abuse within
the first five (5) years from purchase, we will replace it with a new
meter free of charge. For your records, also write the purchase date of
your product here.
Date of purchase:
Note: This warranty applies only to the meter in the original purchase,
and does not apply to the battery supplied with the meter.
Index of Symbol
Consult instructions for use
For
in vitro
diagnostic use only
o
Store between 2-30 C
Contains sufficient for <n> tests
Use by
Lot Number
Control Range
Catalog #
Manufacturer
Authorized Representative
Sterilized using irradiation
Code Number
o
30 C
o
2 C
IVD
LOT
CODE
CTRL
REF
STERILE R
EC
REP
Do not dispose along with household waste
41
42