Manufacturer’s guide and declaration – electromagnetic immunity
This camera is intended for use in the electromagnetic environment specified below.
The user must ensure that it is used in this environment.
Immunity test
IEC 60601
Level of gravity
Compliance
level
Electromagnetic environment - guide
Electrostatic
discharges
IEC 61000-4-2
± 6 kV in Contact
± 8 kV in air
± 6 kV
± 8 kV
The floor must be wooden, concrete or
tiled. If the floor is covered with a synthetic
material, the relative humidity must be at
least 30 %.
Rapid transient
peaks
IEC 61000-4-4
± 2 kV for power
lines
± 1 kV for input/
output lines
± 2 kV
± 1 kV
The quality of the main power supply
must be that of a commercial or typical
hospital environment.
Electric shocks
IEC 61000-4-5
Differential mode
± 1 kV
Shared mode
± 2 kV
± 1 kV
± 2 kV
The quality of the electric power supply
must be that of a commercial or typical
hospital environment.
Power dips,
power failures,
short power
interruptions and
variations in
power voltage
IEC 61000-4-11
• < 5 %
UT
– for 10
ms • 40 %
UT
– for
100 ms • 70
% UT
–
for 500 ms •
< 5 % UT
– for 5 s
< 5 % UT
10 ms
< 40 % UT
100 ms
< 70 % UT
500 ms
5 s
The quality of the main power supply
must be that of a commercial or typical
hospital environment.
If the user of this camera requires it to
continue functioning during interruptions
of the main power supply, it is advisable
for this camera to be powered by an
inverter or a battery.
System
frequency
magnetic field
(50/60 Hz)
IEC 61000-4-8
3 A/m
3 A/m
The system frequency magnetic field
must be at a level that is characteristic of
location in a commercial or typical
hospital environment.