C1
Connector for the Camera head
C2
Main power socket
C3
RS232 output port (connection to Ackermann 16-2033HD-26 and 16-2033HD-32)
C4
Service line
C5
RGBS video output port
C6
DVI-D video output port
C7
Ethernet output port (connection to Master Panel PC)
C8
SDI output port
C9
Connections to Ackermann Fusion devices (light source and insufflator)
C10
Y / C video output port
C11
Composite video output port
C12
Peripheral connections
C13
Equipotential plug
C14
USB storage device port
L1
Touch screen
– 7“ (
± 0.1
)
S1
Standby button
S2
Power button
Z1
Read area for Fusion Card
8 Suggested decontamination procedures
Disinfecting the camera control unit
The camera control unit is a non-heat-resistant medical device and does not withstand
immersion; consequently, it should be disinfected using a non-woven medium saturated with
a disinfectant detergent.
This device is not drip-proof. Do not splash liquid on this product.
Decontamination, methods and/or tools used, is the sole responsibility of the staff
concerned.
The device must always be disconnected before any cleaning procedure is carried out.
This device is not autoclavable.
Existing alkaline solutions for the disinfection of certain medical devices are NOT
RECOMMENDED for disinfecting this device.