Make sure that this certificate is handed over and that a label is attached to the rear side of the device stating the date of
the next inspection.
The manufacturer is relieved of any liability for the operational safety of the device if it is opened, repaired and/or modified
by anyone other than the authorized service personnel.
Medical devices regulations request the user of the device to carry out regular function and safety checks. As manufacturers,
we recommend quarterly inspections which are to be carried out by hospital technicians or other authorized personnel.
You will find a description of the user inspection in the next chapter of these operating instructions.
Before delivery, a complex function and safety test is conducted by the manufacturer for every device. The results of this
quality control are documented in the inspection protocol, which is supplied with the device. The values of the suction and
irrigation performance are valid as guidelines for the user inspection.
12. QUARTERLY INSPEcTIoNS
12.1. CHeCkIng THe pneumaTIC ConTrol (manual ConTrol/Double fooTSWITCH)
For this test you will need:
• The device
• Pneumatic control (manual control/double footswitch)
The pneumatic controller test procedure is carried out in the following steps:
1. Plug the AC plug into an AC outlet and switch on the device using the AC power switch. The STANDBY LED light yellow.
2. Connect the pneumatic control (manual control/double footswitch) to the device. Ensure that the connection is tight
and there are no leaks.
3. Activate the device by pressing the push button ON/OFF.
4. The device performs a self test.
5. Start the device by pressing the push button START/STOP. The START LED light green.
6. Pressing the manual control (blue or red button) I double footswitch (left or right footswitch) sets the associated roller
wheel (SUCTION/IRRIGATION) in motion.
If this does not happen, repeat the procedure. If the second attempt also fails, please contact the relevant authorized
service partner.