Pneumatic double footswitch:
M
Left footswitch- IRRIGATION
N
Control connector - IRRIGATION (blue)
0
Control connector - SUCTION (red)
P
Right footswitch - IRRIGATION
8. funCTIonal buTTonS, STarT-up, funCTIonal TeST & operaTIon
8.1. SeTTIng up
1. Plug the AC plug into an AC outlet and switch on the device using the AC power switch. The STANDBY LED light yellow.
2. Find a suitable location for the double footswitch if required, and connect it to the device. Ensure that the tube is not
kinked or compressed. If you do not require the double footswitch and wish to use the manual control, then connect this.
The tube set must not be in place before the self test or be installed while the self test is in progress!
3. Activate the device by pressing the push button ON/OFF .
4. The device performs a self test.
Self test sequence:
a. All lights come on for 1 sec and then go out again.
b. The left and right bargraph show the software version for 1 sec.
c. 1st segment of the left bargraph display lights: checking the pressure sensor for pneumatic control
d. 2nd-5th segments of the left bargraph light up: checking the irrigation function of the motor
e. 6th-8th segments of the left bargraph light up: checking the suction function of the motor
f. All segments are unlit again. The device signals that is it is ready for operation with a triple beep signal and the ON and
STOP LEOs light.
5. Now install the tube set as described in section 6.4.
6. Carry out a functional test.
8.2. fUNcTIoNAL TEST
Each time the device is used, a functional test must be carried out first.
1 . Start the pump by pressing the push button START/STOP. The START LED light green.
2. Pressing the IRRIGATION button on the pneumatic control causes fluid to be pumped. Bleed the tube set until fluid exits
from the end of the instrument.
3. Pressing the SUCTION button on the pneumatic control causes the suction roller to rotate.
4. The functional test is complete. Stop the pump by pressing the push button START/ STOP. The STOP LED light yellow.