Ackermann
STeRIS® V-PRO™ 1 PLUS
•
Lumen cycle
•
Non lumen cycle
STeRIS® V-PRO™ maX
•
Lumen cycle
•
Non lumen cycle
•
Flexible cycle
STeRIS® V-PRO™ 60
•
Lumen cycle
•
Non lumen cycle
•
Flexible cycle
These methods must only be applied to devices which are specially designed to be compatible with STeRIS®
V-PRO™.
STeRIS® V-PRO™ sterilisation may lead to changes in the outward appearance of devices, which do not
necessarily compromise their functionality.
Devices must be carefully dried prior to loading into the V-PRO sterilisation chamber. Loads containing
moisture could cause the cycle to be cancelled.
ackermann recommends the use of a cleaning agent and sterilisation method to prevent unknown rates
of deterioration of materials due to interactions between the material and chemical products arising from
different cleaning and sterilisation processes.
Use only STeRIS® V-PRO™ compatible instrument trays in the sterilisation chamber. These trays
are specially designed for hydrogen peroxide vapour sterilisation.
Use only polypropylene sterilisation packaging and polyolefine bags approved by the FDa.
Do not use paper bags or sterilisation envelopes containing wood pulp or cotton.
Comment:
The devices judged by ackermann to be compatible with the STeRIS® sterilisation process have been approved
with a minimum of one hundred STeRIS® cycles
.
InstRuCtIons FoR the use oF steRIs
®
v-PRo
™
systems
1. Clean and prepare the devices as recommended in the Cleaning section of this user manual. Make sure the
devices are fully dried.
2. Place the devices on STeRIS® V-PRO instrument trays, wrap them in an FDa-approved polypropylene
sterilisation wrapping or place them in polyolefine bags. Place STeRIS® V-PRO indicator strips in all trays
and bags.
3. Load the instrument tray into the STeRIS® V-PRO™, steriliser, arranging it so that the hydrogen peroxide
vapour can envelop it. Do not allow any item to touch the steriliser wall.
4. Consult the STeRIS® V-PRO™ sterilisation system user manual for detailed instructions.
5. Contact STeRIS® for the most up-to-date information about sterilisation using STeRIS® V-PRO sterilisation
systems
.