3.7
3.8
3.9
3.10
3.11
Ch
ap
te
r
3
The device is a mobile device, and may be moved using a trolley.
movInG the meDICAl DevICe
There are no known undesirable secondary effects associated with the DUO Full HD (control unit and camera
head). Please refer to the user manuals for other devices used during surgical endoscopy or diagnostic
procedures (gynaecology, Laparoscopy, arthroscopy, eNT or spinal endoscopy).
unDesIRABle seConDARy eFFeCts
The device must only be opened by a competent technician approved by the manufacturer.
This device must not be modified without the manufacturer’s authorisation. If the medical device is
modified, a test and inspection must be carried out to ensure that the medical device satisfies the safety
requirements.
Only the fuses below the switch at the rear of the device may be replaced. The power should be switched off
and the fuses checked and replaced if necessary; only T 1.6 a-250 V UR time-delay fuses should be used
(only those marked UL/CSa).
AssemBly AnD DIsAssemBly oF the PRoDuCt
any serious incident concerning the medical device or its accessories, with the exception of the expected
secondary effects must be reported to the relevant competent authorities and to the manufacturer as soon as
possible. generally speaking, the notification period should take into account the seriousness of the incident.
Consult local applicable regulations.
Manufacturer’s contact details: please see page 18 (CONTaCT) of the manual.
vIGIlAnCe
This device is not sterile; only the camera and lens may be sterilised.
When connecting the light cable or camera to the control channel, do not force the connectors, as this may
damage them.
abide by the proper usage and storage conditions (see chapter 9).
ensure there is sufficient airflow to prevent any overheating within the device: at least 15 cm clear all round
the device. Do not cover it, and ensure that all of the device’s feet are present.
If the power cord is damaged, immediately switch off the device. It is dangerous to operate this device with
a damaged cord.
after the source has been used, when removing the fibre from its light guide, the temperature of the metal
part will be very high and could cause burns.
The light intensity at the fibre outlet may cause eye damage. Be careful handling it when the device is in use.
PARtICulAR WARnInGs