System Information and Settings
ACIST RXi System User’s Guide |
55
901700-001,01 2019-09 English
System Settings
Lab ID, Mobile or Stationary
Press
ACIST > System Settings > Lab ID
submenu to view the Lab ID settings
screen, as shown below:
Lab ID
Pop-up
keyboard
Icon
Current
Lab ID
setting
Press
Mobile
or
Stationary
to select the desired mode depending on the
method of installation of the ACIST RXi® System at your facility.
• In Mobile mode, the ACIST RXi System is designated as a mobile
device. In this mode, the system prompts you to zero the aortic
pressure on the ACIST RXi System for every patient case.
• In Stationary mode, the ACIST RXi System is designated as a
stationary device, and is intended to remain installed in one room. In
this mode, zeroing the aortic signal input on the ACIST RXi System is
only required at installation.
WARNING
If the system is set to Stationary mode and you move the system from one lab
to another, you must manually zero the RXi aortic pressure after you zero the
hemodynamic monitoring system or an incorrect FFR value may result.