Other Specifications
98
| ACIST RXi System User’s Guide
901700-001,01 2019-09 English
Technical Specifications
Cable Lengths
Cable
Length
Power cable (power supply to main
power)
8.2 to 10 feet (2.5 to 3 meters)
System console to power supply
5 feet (1.5 meters)
Pressure Signal Input
Signal Type
Specification
Pressure range
-30 to +300 mmHg relative atmosphere thoughout
580 to 800 mmHg absolute atmospheric pressure
Pressure accuracy
±3% of reading or ±3 mmHg of reading over pressure
range, whichever is greater, including the combined
effects of sensitivity, repeatability, non-linearity, and
hysteresis
Pressure drift
<7 mmHg over 1 hour
Frequency response
Response at 10 Hz within 3 dBA of the response at 1 Hz
Input sensitivity
Compatible with 20 mmHg/V or 100 mmHg/V Pa signal
input
Pressure Signal Output
Signal Type
Specification
Output sensitivity
5 uV/V/mmHg nominal scale strain gauge bridge (ANSI/
AAMI BP22)