Start the ACIST RXi® System
30
| ACIST RXi System User’s Guide
901700-001,01 2019-09 English
Basic Operating Procedures
Power On
WARNING
In order to avoid the potential for electric shock to the patient, do not simultaneously
touch the patient and either the system console screen, USB, Ethernet or hemodynamic
connectors located behind the sterile drape.
Start the ACIST RXi System:
1. Ensure the ACIST RXi System has been properly installed and securely
attached on the bed rail or an ACIST CVi® Injector.
2. Turn on the system:
• Facing the RXi console, use the switch on the right side of the unit to
turn on the ACIST RXi System.
ON power
Off power
• The green system status indicator LED located on the console’s front
right side shows that power is on.
Power ON/OFF Switch
System Status Indicator LED
3. Wait approximately 30 seconds for the system to start. The main screen
is displayed when the system is ready for setup. The status line provides
instructions to start the setup procedures. If errors are detected during
startup, the touchscreen displays an error message.