11

12

This way you receive results that are as accurate as possible

A)

 Wait at least 15 minutes after your last sip of alcohol. Do not eat, drink or smoke during that time.

B)

 Do not open the casing.

C)

 Do not keep the breathalyzer in cold, hot, moist or badly ventilated places.

D)

 Keep the breathalyzer out of the reach of children.

E)

 Conduct three inlet tests after long storage before conducting the actual alcohol test.

F)

 Do not drop the breathalyzer.

G)

 Change the battery if it is depleted. Diminished battery capacity leads to incorrect test results.

H)

 Do not conduct tests in loud (sound waves disrupt the sensor), drafty or poorly ventilated places.

I)

 After a test result of 1.000 per mill (per mill = EU setting) / 0.100% BAC (% BAC = US setting) / 

0.500mg/L (mg/L = UK setting) or higher the sensor needs rest time. In this case wait at least 

three minutes before starting the next test.

J)

 Do not conduct tests in very cold, very hot or an unventilated, smoky ambience.

K)

 Store the breathalyzer in a way to avoid penetration by dust or liquids.

L)

 Blowing too hard may cause damage to the sensor as saliva flows into it.

Preparation for the test

1. Install batteries

Push the battery cover (on the back of the device) downwards and open it. Install the 2 AAA 

batteries (delivered together with the device) in the battery compartment while taking note of 

the battery polarity. Close the battery cover again.

2. Insert a mouthpiece

Insert a mouthpiece into the slot on the left front side of the breathalyzer. Ensure whether it securely 

fits or not. It is highly recommended to use an extra mouthpiece for every test for sanitary reasons.

Conducting the measurement

1. 

Press the start button on the front of the alcohol tester. The device beeps 1x and alternately 

displays the number of tests performed so far and the last test result (as long as the start button 

remains pressed). If you no longer press the start button, „WAIT“ will be displayed at the top of the 

display and a countdown will start.

2. 

Wait for the countdown, there are several lines moving on the display, the device warms up. At 

the end of the countdown the device beeps 2x and „blow“ appears on the display. 

3. 

As soon as „blow“ appears on the display, start to blow into the attached mouthpiece (just 

breathe slowly and without interruption, as if you were whistling). The device beeps continuously 

during the breath test.

4. 

Once the beep stops and you hear 2 clack tones, the test is finished. If the breath sample was 

interrupted or too weak, „Flo“ appears on the display. In this case, perform the test again and 

press the start button for 3 seconds. 

5. 

After a successful breath test, the device immediately evaluates the measurement result. The 

digit „0“ moves briefly from the left to the right side of the display and at the end of the evalua-

tion you hear 1 beep sound. The measurement result is finished and then on the display the test 

result is shown with 4 digits (and 3 decimals in the standard measuring unit per mille). After 10 

seconds, you hear 5 beeps and the backlight will turn off (but the measurement result will still be 

displayed). For a new test, simply press the start button for 3 seconds.

Note:

 The device switches off automatically after approx. 3 minutes. You can also switch off the 

device manually by pressing the start button continuously for approx. 5 seconds, then you hear 

a double beep sound and you will see „Out“ on the display (additionally on top right „WAIT“ is 

shown), then the device is switched off.

ACE Instruments recommends

 Always use a new mouthpiece for each person and test that is conducted. ACE Instruments 

assumes no liability for any damages which arise when several people use the same mouthpiece 

for multiple tests.

 Only use the provided single packed and hygienic ACE Instruments mouthpieces. During each 

test the Breathalyzer’s internal and external parts come into contact with human secretion (saliva, 

blood, sweat, etc.)!

 Use safety gloves when pulling out a mouthpiece used by another person! You can buy appro-

priate gloves at ACE Technik (www.ace-technik.de).

IMPORTANT NOTICE

 WHEN YOU HAVE EATEN, DRUNKEN ALCOHOL OR SMOKED A CIGARETTE, WAIT AT LEAST 15 

MINUTES BEFORE STARTING THE MEASUREMENT. ANY TEST CONDUCTED PRIOR TO THE EXPI-

RATION OF THE 

RECOMMENDED WAITING PERIOD WILL RESULT IN INACCURATE RESULTS.

 IF THE TEST RESULT IS BELOW 0.100‰ (PER MILL) OR 0.010% BAC (US setting) OR 0.050mg/L 

(UK settting) THEN THE DEVICE SHOWS 0.00‰ (OR % BAC OR mg/L – depending on the setting) 

ON THE DISPLAY. THIS HAPPENS BECAUSE AT A LOW RANGE OF ALCOHOL CONCENTRATION IT 

IS POSSIBLE TO MEASURE OTHER MATERIALS IN YOUR MOUTH.

 This breathalyzer is a high quality and sensitive measurement device. To maintain the accuracy 

of the sensor, it should regularly be used. If the device is not used for an extended period of time, 

the sensor can dry out. ACE Instruments recommends to use the breathalyzer at least 2 times per 

month (even if you have not been drinking alcohol).

 Due to changes in the product, the procedures stated can change. This can take place without 

prior notice or adaptation of the operating manual and is no cause for complaint, return or price 

reduction. The user is responsible for recognizing discrepancies between the operating manual 

and the actual handling of the breathalyzer and must react accordingly.

Only the usage of original ACE Instruments mouthpieces

 Guarantees test result accuracy

 Guarantees 100% hygenity

 Guarantees full warranty of this device

ACE INSTRUMENTS ADVISES

EVEN IF YOU GET A RESULT MEANING „SAFE RANGE“, YOU SHOULD NEVER DRIVE AFTER 

CONSUMING ALCOHOL OR DO ANY OTHER ACTIVITIES THAT ARE SEVERELY IMPAIRED BY 

ALCOHOL CONSUMPTION AND MIGHT ENDANGER YOUR HEALTH AND SAFETY OR THE HEALTH 

AND SAFETY OF OTHERS.

Summary of Contents for II Basic plus

Page 1: ...omat net Atemalkoholtester Bedienungsanleitung DE Breathalyzer Operating Manual EN Alcotest Notice d utilisation FR Alkohol Tester Gebruiksaanwijzing NL Test alcolico Istruzioni d uso IT Alcohol metro...

Page 2: ...t ck k nnte sich Restalkohol befinden der bei einem erneuten Gebrauch desselben Mund st ckes das neue Testergebnis beeinflusst damit w rde z B bei einer n chternen Person eventuell eine Alkoholisierun...

Page 3: ...asons but also because an already used mouthpiece can influence the next measurement result The mouthpiece could contain residual alcohol which would affect the new test result when the same mouthpiec...

Page 4: ...Erhaltung der Messgenauigkeit die ses Ger tes und der Sensor Garantie empfehlen wir eine regelm ige Kalibrierung Sie k nnen das Ger t zum Kalibrieren an die unter www alkomat net angegebene Service A...

Page 5: ...s zuvor angebrachte Mundst ck zu pusten atmen Sie dazu einfach langsam und ohne Unter brechung aus hnlich als ob Sie pfeifen w rden Das Ger t piepst durchgehend w hrend der Atemprobe 4 Sobald das Piep...

Page 6: ...trag m glich ist Durch nderungen am Produkt k nnen sich die beschriebenen Abl ufe ndern Dies kann ohne Vorank ndigung und Anpassung der Bedienungsanleitung geschehen und ist kein Grund zur Reklamation...

Page 7: ...der Anwender den Kostenvoranschlag ab sendet ACE Instruments das Produkt unrepariert unkalibriert zur ck und das Produkt wird entsprechend gekennzeichnet Sollte das Produkt diese Gew hrleistung nicht...

Page 8: ...n service you can send the device to the service address specified at www alkomat net The device must only be used pursuant to the instructions mentioned in this manual All measured values determined...

Page 9: ...d at the end of the evalua tion you hear 1 beep sound The measurement result is finished and then on the display the test result is shown with 4 digits and 3 decimals in the standard measuring unit pe...

Page 10: ...e for complaint return or price reduction The user is responsible for recognizing discrepancies between the operating manual and the actual handling of this Breathalyzer and must react accordingly Bat...

Page 11: ...estimate ACE Instruments will send the product back unrepaired not calibrated and the product will be labeled accordingly Should the product not fulfill the warranty the claim of the buyer and the obl...

Page 12: ...nvoyer l appareil pour l talonnage l adresse de service indiqu e sur www alkomat net L appareil ne doit tre utilis que dans le complet respect des instructions d utilisation Toutes les valeurs mesur e...

Page 13: ...la mesure Le chiffre 0 se d place bri vement du bord gauche vers le bord droit de l affichage et la fin de l valuation un bip sonore retentit Le r sultat de la mesure est ensuite affich sur l cran en...

Page 14: ...n un motif de r clamation de retour ou de r duction du prix Il est de la responsabilit de l utilisateur d identifier les divergences entre le mode d emploi et l utilisation effective et de r agir en c...

Page 15: ...renvoie le produit non r par non calibr et le produit est marqu en cons quence Si le produit ne remplit pas cette garantie le droit de l acqu reur et les devoirs de ACE Instruments au gr de ACE Instru...

Page 16: ...cijfers met 3 decimalen en zijn interpreteerbaar 1 1 het display geeft bijvoorbeeld de waarde 0 36 0 36 per duizend De sensor ingebouwd in de alcoholmeter kan worden gekalibreerd Om de meetnauwkeurigh...

Page 17: ...geval de test opnieuw uit en druk op de startknop gedurende 3 seconden 5 Na een succesvolle ademtest evalueert het apparaat onmiddellijk het meetresultaat Het cijfer 0 gaat kortstondig van de linker...

Page 18: ...een klacht retour of korting Het is de verantwoordelijkheid van de gebruiker om verschillen tussen de gebruiksaanwijzing en de eigenlijke bediening op te merken en dienovereenkomstig te handelen De ba...

Page 19: ...ker eventueel een prijsopgave Wijst hij deze van de hand zendt de ACE Instruments het product niet gerepareerd niet gekalibreerd terug en op het product wordt dit ook vermeld Mocht het product deze wa...

Page 20: ...sura di questo dispositivo e la garanzia del sensore si consiglia una calibrazione regolare possibile inviare il dispositivo all indirizzo di servizio specificato su www alkomat net Questo prodotto pu...

Page 21: ...do destro del display e alla fine della valutazione viene emesso 1 segnale acustico Il risultato della misurazione viene visualizzato sul display con 4 cifre con 3 cifre decimali nell unit di misura s...

Page 22: ...nuto a riconoscere le differenze tra le istruzioni per l uso e l uso effettivo e a reagire di conseguenza Note alle batterie Il dispositivo a cui si riferiscono queste istruzioni per l uso contiene 2...

Page 23: ...esa Se l utente rifiuter il preventivo di spesa ACE Instruments restituir il prodotto non riparato non calibrato ed il prodotto verr contrassegnato di conseguenza Se il prodotto non dovesse adempire a...

Page 24: ...es decir si se muestra un valor de 0 36 eso quiere decir que el valor registrado es de 0 36 g de alcohol kg de sangre El sensor integrado en el alcohol metro se puede calibrar Para mantener la precis...

Page 25: ...y se escuche un tono de doble clic la prueba habr finalizado Si la muestra de aliento se interrumpi o fue demasiado d bil aparece Flo en la pantalla En este caso vuelva a realizar la prueba y presione...

Page 26: ...reducci n del precio Es responsabilidad del usuario reconocer y reaccionar a las divergencias entre el manual de instrucciones y el funcionamiento real Notas para las bater as El aparato al que se ref...

Page 27: ...icho producto ser identificado como tal En caso de que el producto no cumpliera con las condiciones de esta garant a las exigencias del comprador y las obligaciones de ACE Instruments se limitar n exc...

Page 28: ...49 50 Notizen Notes...

Page 29: ...51 ACE Handels Entwicklungs GmbH Staufenstra e 1 Hallen 8 14 83395 Freilassing Deutschland Germany Tel 49 0 8654 779 53 31 Fax 49 0 8654 779 96 94 info alkomat net www alkomat net...

Reviews: