When looking at the front of the Newport cabinet, Nitrous Oxide is on the left as
shown by the blue Nitrous Oxide gauge. Oxygen is on the right as shown by the
oxygen gauge.
Opening Newport Cabinet
Open the Newport cabinet by placing one hand on the front cover and the other in
the handle on the side. Pull handle back all the way.
Install Gas Cylinders
1. Undo Velcro straps
2. Angle tank into position
3. Strap tanks loosely
4. Remove regulators from accessories box
5. Attach regulators to tanks by loosening yoke handle, verifying seal is in
place, aligning seal with gas opening, aligning pins with holes in yoke,
tightening yoke onto cylinder
NOTE: Regulators are pinned to match gas cylinders. Connect gas hoses to
regulator outputs.
NOTE: Hoses and regulators use DISS fittings to prevent mix up. Rotate cylinders
so that regulators fit inside.
Open Gas Cylinders
Pick one tank of each gas to be “in use“ tank and attach “in use” label to cylinder.
Open these cylinders using the cylinder wrench provided in accessory box. Close
Newport cabinet.
Scavenging Circuit
- Install bag tee from accessories box into front of Newport.
- Attach scavenging circuit to bag tee.
- Connect vacuum hose to system vacuum.
- Install nasal hood into scavenging circuit and start vacuum flow per
scavenging instructions.
Newport Setup and Connections