PC UPRIGHT 2-CYLINDER PORTABLE
2-CYLINDER PORTABLE
SYSTEM ASSEMBLY
1. Mount the flowmeter to the portable manifold:
a. Remove hole cap from the upper hole (use
sharp edge knife or screwdriver) at rear of
manifold and release set screw (use 5/32"
Allen Wrench) enough to remove flowmeter
mount.
b. Remove protective cap from threads of
mount and remove from manifold. Thread
flowmeter mount into hole on the bottom of
flowmeter by hand and tighten firmly using
5/32" Allen Wrench placed in hole on side
of flowmeter mount.
c. Place flowmeter on manifold with mount in
pocket.
d. Tighten set screw firmly from rear while
keeping flowmeter oriented toward front of
manifold.
e. Place hole cap into hole behind the set screw
and push it all way in.
2. Attach the hoses to the flowmeter:
a. Run the nitrous oxide hose under the mani-
fold and attach it to the nitrous oxide DISS
fitting located on the back of the flowmeter.
b. Run the oxygen hose under the manifold
and attach it to the oxygen DISS fitting
located on the back of the flowmeter.
Note: To avoid gas leaks, make certain all hose
attachments are securely tightened
.
3. Connect the gas cylinders to the manifold’s
“E” yokes:
a. Connect the nitrous oxide gas cylinders to
the “E” yoke on the N
2
O side
b. Connect the oxygen gas cylinders to the
“E” yoke on the O
2
side.
c. It is recommended to use new gasket every
time tank is replaced or removed.
4. (Keep wrench in handy location. It is used to
open and close tanks).
a. Open both gas cylinders and turn off
flowmeter.
b. Check hose connections for leaks using
soap solution.
c. Tighten fittings as needed to stop leaks.
5. Verify flowmeter operations.
Note: See page 23 for attaching rubber goods,
including scavenging circuit.
Flowmeter
Mount
(make sure lock
washer is still in
position prior to
installation into
flowmeter)
Manifold
Set Screw
Hole Cap
O
2
Hose
Stand Post
N
2
O Hose
E-Yoke
(1 of 2)
Mobile Base
Locking Washer
(remove protective
cap from flowmeter
mount prior to
installation into
flowmeter)
Identify the following
components in the 2-cylinder
portable system package:
• Flowmeter
• Portable Manifold (with
DISS Hoses and mobile
stand
attached)
• E Cylinder Wrench (with
chain attached)
14 5/8"
17 1/4"
39 3/8"
O
2
Resuscita
tor
200
400
600
800
1000
200
400
600
800
1000
4000
3000
2000
1000
-7-