Accutron Ultra PC % User Manual Download Page 12

12

O

2

Resuscitator

Oxygen Flush

Mixed Gas Outlet
(3/8

"

) Gas Specific Instant Fitting

Nitrous Oxide Inlet
(1/4

"

) Gas Specific Instant Fitting

Mounting
Post With
Wingnut

Serial #
Label

Oxygen Inlet
(5/16

"

) Gas 

Specific Instant
Fitting

% N

2

O

Knob

N

2

O

Flow
Tube

O

2

Flow
Tube

Flow
Control
Knob

II. PC CABINET MOUNT UNIT

(For Central Gas Systems)

4 5/8"

(12,2cm)

9 3/4" (25cm)

4 1/4"

(11cm)

1 5/16"

(3,3cm)

2 5/16"
(5,8cm)

4"

(10cm)

6 1/4" (15,9cm) MAX

8 1/2" (21,6cm)

4 1/2"

(11,4cm)

MAX

1/4"

(,64cm)

HOLES 2x

R 1/4" (,64cm) MAX

TECHNICAL SPECIFICATIONS

Physical 5" x 10" x 4" (13 x 25 x 10 cm)
Weight

5.9 lbs

(2, 7 kg)

Gas Supply Requirement

Oxygen: 50-55 PSI at 100 LPM
N

2

O: 50-55 PSI at 10 LPM

Gas Fittings

Oxygen Riser: Male DISS (CGA 1240) 

with check valve

N

2

O Riser: Male DISS (CGA 1040) 

with check valve

CM Fittings

Oxygen Inlet: 5/16" instant fitting
N

2

O Inlet: 1/4" instant fitting

Mixed Gas Out: 3/8" instant fitting
Oxygen Resuscitator: 1/4" I.D. 

Quick Disconnect

Gas Delivery

Oxygen Flush: Min 20 LPM
Oxygen Flow: 1 - 10 LPM

(Accuracy, +/- 0.5 LPM)

Oxygen Resuscitator: Min 100 LPM

100% O

2

Nitrous Oxide Flow: 0 - 7 LPM 
(Ac/- 0.5 LPM)
Nitrous Oxide %: 0-70%*

*50% Max Units Available

Note: Cabinet Mount units require remote outlet.

Summary of Contents for Ultra PC %

Page 1: ...1 Ultra PC Flowmeter User Manual...

Page 2: ...tients Pediatr Dent 2009 31 6 148 151 Available at http www aapd org media policies asp Accessed February 23 2010 WARNINGS To be used only by a professional trained in the use of nitrous oxide Patient...

Page 3: ...performance and will void the warranty Safety features contained in this notice should be routinely checked to assure proper function If any of these safety features are not functioning properly conta...

Page 4: ...isponible sur le site Web http www aapd org media policies asp Acc s le 23 f vrier 2010 AVERTISSEMENTS Doit tre utilis uniquement par un professionnel form dans l utilisation de l oxyde azoteux Le pat...

Page 5: ...on rendement et annulera la garantie Vous devez v rifier couramment les caract ristiques de s curit de cette brochure pour assurer un bon fonctionnement Si l une des caract ristiques de s curit ne fon...

Page 6: ...6...

Page 7: ...on 13 III Newport Portable System 14 IV Safety Features 18 V General Instructions 18 VI Directions for Use 20 VII Periodic Equipment Checks 21 VIII Rubber Goods Attachment 23 IX Emergency Oxygen Equip...

Page 8: ...n Inlet Male DISS CGA 124 N2O Inlet Male DISS CGA 1040 Mixed Gas Outlet O D 875 2 2 Oxygen Resuscitator 1 4 I D Quick Disconnect Gas Delivery Oxygen Flush Min 20 LPM Oxygen Flow 1 10 LPM Accuracy 0 5...

Page 9: ...UPRIGHT CENTRAL GAS SYSTEM MOUNTING OPTIONS Mobile Stand Bottom Mount Under Cabinet Slide Back Mount Wall Arm Mounts Back Mount 2 Gas Supply Lines Connect To Wall Outlet Mobile Stand 1 2 20 Threaded...

Page 10: ...hose attachments are securely tightened 3 Connect the gas cylinders to the manifold s E yokes a Connect the nitrous oxide gas cylinders to the E yoke on the N2O side b Connect the oxygen gas cylinders...

Page 11: ...a Run the nitrous oxide hose under the mani fold and attach it to the nitrous oxide DISS fitting located on the back of the flowmeter b Run the oxygen hose under the manifold and attach it to the oxyg...

Page 12: ...AX TECHNICAL SPECIFICATIONS Physical 5 x 10 x 4 13 x 25 x 10 cm Weight 5 9 lbs 2 7 kg Gas Supply Requirement Oxygen 50 55 PSI at 100 LPM N2O 50 55 PSI at 10 LPM Gas Fittings Oxygen Riser Male DISS CGA...

Page 13: ...alve Risers Ultra PC Flowmeter 30 25 20 1 5 1 0 5 0 inHg VACUUM a Standard Bag Tee Contains mixed gas outlet b Remote Flow System RFS Contains mixed gas outlet and vacuum controller with gauge display...

Page 14: ...Tank Restraint Gas Cylinders Line Pressure Gauges Removable Bag Tee O2 Resuscitator Fitting Regulators and E Yokes Handle Warning Do not hold handle when opening cabinet III Newport 4 Cylinder Portab...

Page 15: ...ing seal with gas opening aligning pins with holes in yoke tightening yoke onto cylinder NOTE Regulators are pinned to match gas cylinders Connect gas hoses to regulator outputs NOTE Hoses and regulat...

Page 16: ...e front panel of the Newport The Newport has features incorporated to make cylinder change out easy To assure proper operation of the system 4 cylinders must be properly installed at all times Please...

Page 17: ...17 Flowmeter is behind shield at front of unit Serial Label Newport Accessing Serial Number...

Page 18: ...ing with Accutron Complete the warranty card supplied with flowmeter Ultra and mail to Accutron Completion of this step ensures proper device warranty coverage Accutron analgesia machines carry a two...

Page 19: ...at 2 Tablets 70369 1 69781 MN 01 Note Reference the disinfectant s label instructions for appropriate application procedures and inactivation of specific organisms Care should be taken not to let disi...

Page 20: ...ollows Turn the Flow Knob in a counter clockwise direction to increase flow rate and in a clockwise direction to decrease flow rate Total flow rate is the sum of the N2O flow rate and the O2 flow rate...

Page 21: ...rvoir bag is filled in 5 10 seconds Release blockage of oxygen flow from bag tee outspout after the 5 10 seconds required to conduct the test has elapsed Quick Reference Chart for Calculating N2O N2O...

Page 22: ...t 1 Flowmeter should be in the off position 2 Reservoir bag should be connected to the downspout of the bag tee 3 Corrugated tube should be connected to the outspout of the bag tee but not connected t...

Page 23: ...on flowmeters are equipped with a resuscitation connector which allows for the attachment of an oxygen demand valve See Accutron Catalog Emergency Oxygen Equip ment Section for information regarding A...

Page 24: ...nk 2 Replace with full cylinders 3 Check O2 gas line pressure Be certain O2 pressure is 50 55 psi If not call dealer for service Check gas line pressure Be certain line pressure is 50 55 psi If not ca...

Page 25: ...of this agreement or the products As used in this paragraph consequential damages include but are not lim ited to lost profits lost revenues property damage personal injury damage to the Purchaser or...

Page 26: ...horized Accutron distributor Contact distributor where unit was purchased Check Troubleshooting Guide on page 24 prior to contacting distributor XIV ASSISTANCE For Assistance contact your local dental...

Page 27: ...treet Address ____________________________________________________________ City State Zip ____________________________________________________________ Accutron Flowmeter Unit Model o Ultra PC o Ultra...

Page 28: ...28 1733 W Parkside Lane Phoenix Arizona 85027 Toll free 800 531 2221 Local Phone 623 780 2020 Fax 623 780 0444 www accutron inc com...

Reviews: