Accutome PachPen Manual
Safety Precautions
There are several areas in the use of the PachPen that require special attention, as
they may pose a safety threat.
The PachPen has an enclosure rated Degree of Protection of IP32. The enclosure
provides protection for objects larger than 2.5 mm and dripping water. In the event
of a spill contacting the unit, wipe the unit completely dry before returning it to
service.
Disinfection and Cleaning
Disinfection issues are confined to the Accutome probe that comes in contact
with the patient’s eye. In order to prevent the transmission of disease, refer to the
OSHA and CDC guidelines for proper control of disinfection. These guidelines are
frequently updated so be sure to contact OSHA, CDC, or your local disease control
agency for the latest information and disinfection technique.
The probe must be cleaned between patients to prevent patient-to-patient transfer
of infection. It is the user’s responsibility to ensure that the relevant standards are
maintained and that the products and procedures are effective and appropriate
for ophthalmic applications. The following information is provided for the guidance
of users, and specific products are mentioned for illustration only. Accutome does
not endorse the use of these or any other product. Products must be used in
accordance with the manufacturer’s instructions.
Warnings
DO NOT AUTOCLAVE THE PROBES OR CONNECTORS.
DO NOT IMMERSE THE PROBE’S CABLES OR METAL CONNECTORS. ALLOW TO DRY
BEFORE USE.
DO NOT IMMERSE THE PROBE TIPS IN TAP WATER. USE DISTILLED WATER FOR
CLEANING AND DISINFECTION
Cleaning Procedure at Point of Use
1. Wear protective gloves when performing the cleaning process.
2. Use a soft cloth lightly dampened in a mild soap or compatible cleaning
solution to remove any particulate matter or body fluids that remain on the
probe or cable.
3. To remove remaining particulates, rinse with a distilled water —dampened
cloth to remove soap residue, and then wipe with a dry cloth.