TRA 

Direct input needs to load the value 

of calibration weight,Load the 

calibration weight,press 

[ALARM],Exit after completion of 

calibration. 

FUN06-USR

   

User manual 

calibration 

TRA 

3.5 

TRA 

WEIGHT 

window 

UNIT WEIGHT 

  window 

FUN07-LIT

 

Backlight mode 

setting 

WEIGHT   

window 

UNIT WEIGHT 

window 

FUN08-BUD

 

Baud rate Setting 

(this function is 

temporarily 

unavailable) 

LIT 

Au         

BUD 

1200 

LIT 

On         

BUD 

2400 

LIT 

Off       

BUD 

4800~19200 

WEIGHT 

  window 

UNIT WEIGHT 

  window 

FUN09-ODE

 

Communication 

format setting 

[this function is 

temporarily 

unavailable] 

WEIGHT   

window 

UNIT WEIGHT 

  window 

FUN10-UPT

 

AUTO setting 

ODE 

UPT 

On         

ODE 

UPT 

On         

ODE 

WEIGHT 

  window 

UNIT WEIGHT   

window 

FUN11-ALR

   

Alarm setting 

ODE 

ALR 

Cnt     

 

 

 

ALR 

Elt     

 

 

Description: in the above steps, press [warning preset] to enter the sub-menu, press [quantity setting] to 

select, press [warning preset] to save and exit, and then press [quantity setting] to enter the next step. 

   

 

 

 

 

10. Quick  calibration  by  users 

(not  applicable  for  the  initial  calibration  of  replacement 

mainboard or sensor)

 

 

When starting up self-check 0~9 hop number, quickly enter the user's calibration password [9158]. 

 

WEIGHT 

UNIT WEIGHT 

QUANTITY 

0.0   

put 

Au-cal 

After loading the calibration weights above the half-range and stabilizing for 3 seconds,   

the automatic calibration completes and exits 

 

 

 

 

 

 

 

 

Summary of Contents for DSK Series

Page 1: ...1 User Manual DSK series Scale EN ...

Page 2: ...2 ...

Page 3: ...weigh within the specified range when weighing 6 Shall not be used in bad environment 7 When taking the electronic scale do not force on the bracket or weighing plate 8 The maintenance of electronic scale must be maintained by authorized professional personnel 2 Packing list 1 2 3 4 5 Electronic scale Scale plate Power cord Operating instruction Certificate of quality 3 Technical parameters Rated ...

Page 4: ... the scale plate 4 When used the center of gravity of the weighing object must be located in the center of the scale plate and the weighing object should not exceed the scope of the scale plate to ensure its accuracy 5 Before using the electronic balance please heat up the engine for 15 20 minutes Maximum capacity g 3000 6000 15000 30000 Minimum capacity g 20 40 100 200 verification scale interval...

Page 5: ... the sample is finished the buzzer will make a sound and the display screen will display the set number and average single volume 2 U W SET Input a known sample single weight through the keyboard press U W SET when the number window flashes and then place the sample to be counted on the scale plate 3 ALARM By entering the number of alarm upper limit press the ALARM to light up the PSET sign when t...

Page 6: ...n entering quantity or single error to clear and re enter immediately or wait 5 seconds to enter again the previous error will be automatically updated 7 2 Indication sign description UDSP Sample number less than 10 PCS the more sample number the better count UDWT Count sample single weight less than 1 5e show 1 5 times the scale value AUTO In the process of loading counting samples the average we...

Page 7: ...ing WEIGHT window UNIT WEIGHT window FUN04 CAL Factory linear calibration CAP 0 CAL 1 Enter the maximum weight directly through the numeric key CAL 2 WEIGHT window UNITWEIGHT window FUN05 TRA Zero tracking setting CAL 3 TRA 0 CAL 4 TRA 0 5 CAL 5 TRA 0 1 WEIGHT window UNIT WEIGHT window QUANTITY window TRA 1 5 0 PUT AU CAL TRA 2 Add the maximum weight to calibrate the weight and stabilize it for 3 ...

Page 8: ...indow UNIT WEIGHT window FUN10 UPT AUTO setting ODE 0 UPT On ODE 1 UPT On ODE 2 WEIGHT window UNIT WEIGHT window FUN11 ALR Alarm setting ODE 3 ALR Cnt ALR Elt Description in the above steps press warning preset to enter the sub menu press quantity setting to select press warning preset to save and exit and then press quantity setting to enter the next step 10 Quick calibration by users not applica...

Page 9: ...iate environment Accuracy is poor Improper calibration Calibrate Can t calibration Unstable environment Move the electronic counter to the appropriate environment The backlight doesn t work Mode off Enter the menu Settings activation mode Alarm is invalid Mode off Enter the menu Settings activation mode Battery symbol flicker Low battery Charge and replace the battery in time Battery failure Batte...