8
TRA
3
Direct input needs to load the value
of calibration weight,Load the
calibration weight,press
[ALARM],Exit after completion of
calibration.
FUN06-USR
User manual
calibration
TRA
3.5
TRA
4
WEIGHT
window
UNIT WEIGHT
window
FUN07-LIT
Backlight mode
setting
WEIGHT
window
UNIT WEIGHT
window
FUN08-BUD
Baud rate Setting
(this function is
temporarily
unavailable)
LIT
Au
BUD
1200
LIT
On
BUD
2400
LIT
Off
BUD
4800~19200
WEIGHT
window
UNIT WEIGHT
window
FUN09-ODE
Communication
format setting
[this function is
temporarily
unavailable]
WEIGHT
window
UNIT WEIGHT
window
FUN10-UPT
AUTO setting
ODE
0
UPT
On
ODE
1
UPT
On
ODE
2
WEIGHT
window
UNIT WEIGHT
window
FUN11-ALR
Alarm setting
ODE
3
ALR
Cnt
ALR
Elt
Description: in the above steps, press [warning preset] to enter the sub-menu, press [quantity setting] to
select, press [warning preset] to save and exit, and then press [quantity setting] to enter the next step.
10. Quick calibration by users
(not applicable for the initial calibration of replacement
mainboard or sensor)
When starting up self-check 0~9 hop number, quickly enter the user's calibration password [9158].
WEIGHT
UNIT WEIGHT
QUANTITY
0.0
put
Au-cal
After loading the calibration weights above the half-range and stabilizing for 3 seconds,
the automatic calibration completes and exits