- 11 -
lines
Surge
IEC 61000-4-5
±1 kV differential
mode
±2 kV common
mode
±1 kV differential
mode
±2 kV common
mode
Mains power quality should be that of
a typical commercial or hospital
environment.
Voltage
drops,
dips,
and
fluctuations
of
input
power
supply line IEC
61000-4-11
<5 %
U
T
(>95 % dip in
U
T)
for 0,5 cycle
40 %
U
T
(60 % dip in
U
T)
for 5 cycles
70 %
U
T
(30 % dip in
U
T)
for 25 cycles
<5 %
U
T
(>95 % dip in
U
T)
for 5 sec
<5 %
U
T
(>95 % dip in
U
T)
for 0,5 cycle
40 %
U
T
(60 % dip in
U
T)
for 5 cycles
70 %
U
T
(30 % dip in
U
T)
for 25 cycles
<5 %
U
T
(>95 % dip in
U
T)
for 5 sec
Mains power quality should be that of
a typical commercial or hospital
environment. If the user of the
ACCUNIQ BP250 requires continued
operation
during
power
mains
interruptions, it is recommended that
the ACCUNIQ BP250 be powered
from an uninterruptible power supply
or a battery.
Magnetic field of
commercial
frequency
(50/60Hz)
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields
should be at levels characteristic of a
typical
location
in
a
typical
commercial or hospital environment.
Note
U
T is the a.c. mains voltage prior to application of the test level.