AM8040 Precision Thermometer User’s Guide
Rev. 202206V1
25/40
6)
Once the file has finished copying, double click on the icon
“CoolDragon_Air_Reader” to begin installation. The following
screen will display:
7)
“The Base Status” and “Remote” should display “Wireless Base RF
OK” and “Wireless Remote OK” to confirm the wireless connection.
The value after “ADDR” shows the wireless address of AM8040
precision thermometer. The data after “RSSI” is the wireless signal
strength which should be a two-digit number after clicking on
“Start”. The RSSI should be greater than -99dBm. Otherwise, the
wireless signal is too weak and the link is not stable. To increase the
wireless power, users can move the instrument closer to the
computer or increase the height of the Wireless Transceiver. If it is
still too weak, the environment is not suitable for wireless
communication.
8)
To begin wireless data acquisition, click on “Start”. If the wireless
signal is good enough, AM8040 should display “ZB-xx” where “xx”