Thermometer/Alarm (TMD-01) 

Your unit is equipped with a NIST calibrated Thermometer/Alarm that simultaneously displays 
minimum, maximum, and current interior temperatures and provides a visual and audible signal 
when the temperature inside the unit falls outside the set temperature range.  
 

NOTE ABOUT CALIBRATION:

 All units ship with a two-year certificate of calibration in the 

alarm/thermometer box that comes inside your unit. Keep this certificate stored safely for your 
records. Accucold offers re-calibration services through their ISO 17025:2017 certified calibration 
laboratory. It is recommended to have your probe re-calibrated every two years (or one year, 
depending on your state’s health department guidelines). For more information on how to have your 
device re-calibrated, contact us at 

[email protected]

. 

 
The temperature sensor, or probe, is enclosed in a glycol-filled bottle that you insert into the location 
to be monitored once the unit has been programmed. The thermometer/alarm itself can be attached 
to the door of the unit with the magnet on the back. 
 

NOTE:

 A slight variation in temperature is often seen from one thermometer reading to another, 

even when the thermostat is set at a particular temperature. This is normal. If the 
thermometer reading does not fluctuate at all over several readings, temporarily remove the 
thermometer from the storage unit and place it outside the unit at room temperature. Check 
whether the temperature reading rises. If no change in the temperature reading occurs, the 
thermometer is faulty and needs to be replaced.  

 

Programming the Alarms 

The thermometer displays the internal temperature of the unit being monitored, as well as the 
maximum and minimum temperatures reached. The maximum and minimum readings reflect the 
highest and lowest temperatures since the battery was inserted or since the MODE/RESET key was 
last pressed. If the temperature rises above or falls below the set temperature range, the alarm will 
sound. By default, the high alarm temperature setting is -15°C (5°F) and the low alarm temperature 
setting is -50°C (-58°F). To reset these limits, follow these instructions: 
 
• 

High Alarm Temperature Setting:

 Press the HIGH ALARM key ONCE. HI-SET will appear on 

the display. Use the 

▲ 

and 

▼ keys 

to increase or decrease the value. Press the HIGH ALARM 

key again to confirm the setting. The display will return to the current temperature. 
 

• 

Low Alarm Temperature Setting:

 Press the LOW ALARM key ONCE. LO-SET will appear on 

the display. Use the 

▲ 

and 

▼ keys 

to increase or decrease the value. Press the LOW ALARM 

key again to confirm the setting. The display will return to the current temperature. 
 

By default, the low and high alarms should be audible. To turn alarms on/off: 
 

• 

Press any key ONCE to stop the beeping sound. 

The beeping sound lasts for 60 seconds, but 

the high and low indicators will remain on display even when the unit gets back in range. 
Pressing any key will silence the beeping sound, but it will be reactivated every 30 minutes if 
the alarm event still exists.   

•  Press the 

 and 

 keys simultaneously to clear the LO-ALARM/HI-ALARM indicators. 

The alarm will not clear if the alarm event still exists. 

 

For further information about how your alarm work, please read the User Instructions, which come 
packaged with your alarm. 
 

Summary of Contents for VT133

Page 1: ...3 VLT74 VT103 VLT104 Owner s Manual Mode d emploi BEFORE USE PLEASE READ AND FOLLOW ALL SAFETY RULES AND OPERATING INSTRUCTIONS AVANT UTILISATION S IL VOUS PLA T LIRE ET SUIVRE LES R GLES DE S CURIT E...

Page 2: ...uct uses eco friendly hydrocarbon refrigerant and fully complies with California CARB regulations However we are required by California Law to provide the following disclosure statement in every produ...

Page 3: ...ity of Isobutane R600a These are environmentally but flammable They do not damage the ozone layer nor do they increase the greenhouse effect Care must be taken during transportation and setting up the...

Page 4: ...such as wind rain water spray or sunlight The chest freezer should not be located next to ovens grills or other sources of high heat The chest freezer must be installed with all electrical connections...

Page 5: ...bient temperatures may also cause the freezer not to perform properly Avoid locating the freezer in moist areas Plug the freezer into an exclusive properly installed grounded wall outlet Do not under...

Page 6: ...inside the unit as opposed to the air temperature itself As a result these readings may not always display the same temperature reading To Change the Set Point While the set point value is blinking pr...

Page 7: ...ter is faulty and needs to be replaced Programming the Alarms The thermometer displays the internal temperature of the unit being monitored as well as the maximum and minimum temperatures reached The...

Page 8: ...er does not overflow 6 Clean and thoroughly dry the interior of the freezer 7 Replace the drainage plug on the inside floor 8 Close the door and plug in the freezer 9 Reset the thermostat to the desir...

Page 9: ...n freezer door unnecessarily if freezer is off for several hours If a power failure occurs frozen items will stay frozen for at least 24 hours if the freezer is kept closed If the power failure contin...

Page 10: ...warm or hot items might have been stored recently Wait until the freezer has had a chance to reach its selected temperature The freezer has recently been disconnected for a period of time Four hours a...

Page 11: ...stallation installation not in accordance with electrical or plumbing codes or use of products not approved by warrantor 4 Replacement parts or repair labor costs for units operated outside the United...

Page 12: ...les Ils n endommagent pas la couche d ozone et n augmentent pas l effet de serre Lors du transport et de l installation de l appareil il faut veiller ce qu aucune pi ce du syst me de refroidissement n...

Page 13: ...areil l avant de l appareil doit tre compl tement d gag Choisissez un endroit bien ventil avec des temp ratures sup rieures 16 C 60 F et inf rieures 32 C 90 F L appareil ne doit pas tre situ c t de fo...

Page 14: ...ifornia CARB Cependant la loi californienne nous oblige fournir la d claration de divulgation suivante dans chaque produit vendu en Californie L utilisation de cet quipement est interdite en Californi...

Page 15: ...onner correctement viter de placer le cong lateur dans des zones humides Branchez le cong lateur dans une prise murale exclusive correctement install e et mise la terre Ne pas en aucun cas r duire ou...

Page 16: ...ompartiment n est pas refroidi Le voyant rouge marqu Alarm s allume galement lorsque l appareil est branch et reste allum jusqu ce que la temp rature int rieure tombe en dessous de 11 C 12 2 F Comme a...

Page 17: ...e permettra d acc l rer le processus En outre verser de l eau chaude non bouillante sur la glace gel va acc l rer la fusion Si vous devez utiliser un grattoir il ne devrait pas tre en bois ou en m tal...

Page 18: ...aisser la porte entrouverte et bloquer la ouverte si n cessaire ou avoir la porte enlev e Soyez extr mement prudent dans le cas des enfants L unit ne doit pas tre accessible aux enfants jouer Panne le...

Page 19: ...ajout e au cong lateur La porte est ouverte trop souvent La porte n est pas compl tement ferm e Le contr le de la temp rature n est pas correctement r gl Le joint de la porte ne se scelle pas correcte...

Page 20: ...g lateur Le cong lateur touche le mur D placez le du mur et remettre niveau le cong lateur Voir Instructions d installation La porte ne se ferme pas correctement Le cong lateur ne se trouve pas sur un...

Page 21: ...Unis 5 R parations de pi ces ou de syst mes r sultant de modifications non autoris es apport es l appareil 6 Le retrait et la r installation de votre appareil s il est install dans un endroit inaccess...

Reviews: