18
(50-104°F).
•
Make sure that the test strip vial and the control solution bottle have been tightly
capped.
•
Make sure the test strip was used immediately after removing from the test strip vial
or foil pouch.
•
Make sure the control solution was mixed well.
•
Confirm that you are using Accu-Tell brand control solution.
•
Make sure that you followed the test procedure correctly.
After checking all of the conditions listed above, repeat the control solution test with
a new test strip. If your results still fall out of the range indicated on the test strip
vial label or on the foil pouch, your meter or test strips may not be working properly.
DO NOT
use the system to test blood. Contact your dealer for help.
To turn your meter off, just remove the test strip. Dispose of the used test strips as
medical waste. Control results will be not included in your blood glucose averages.
Using the Meter Memory
Your meter automatically stores up to 300 results with the time and date. Test
results are stored from the newest to the oldest. The meter will also calculate the
average values of records from the last 7, 14 and 30 days.
Notes:
•
If there are already 300 records in memory, the oldest record will be erased to make
room for a new one.
•
It is very important to set the correct time and date in the meter, please make sure
the time and date are correct after you change your battery.
•
Control results are not included in the 7, 14 and 30 day average calculation.
Viewing Your Test Results
When your meter is off, press
to turn meter on after a beep, a symbol of strip
flashes on the display, press
again, the 7 day average will appear in the center
of the display with the number of results shown at the top of the display. If you want
to review the memory after you immediately performed a test, with the test result
on the display, press
to see the 7 day average.
Summary of Contents for VGM06