Questionable or Inconsistent Results:
If your pet’s blood glucose result does not match how you feel, please
:
•
Check the expiration date and the discard date of the test strip. Make sure
that the test strip vial has not been opened for more than 6 months.
•
Confirm the temperature in which you are testing is between 5 and 45°C
(41-113°F).
•
Make sure that the test strip vial has been tightly capped.
•
Make sure the test strip has been stored in cool, dry place.
•
Make sure the test strip was used immediately after removing from the test
strip vial or foil pouch.
•
Make sure that you followed the test procedure correctly.
•
Perform a control solution test (See
Performing a Control Test
for
instructions).
After checking all of the conditions listed above, repeat the test with a new test strip.
If you are still unsure of the problem, please contact your local dealer.
Removing the Used Lancet
Unscrew the lancing device cover. Place the safety tab of the lancet on a hard
surface and carefully insert the lancet needle into the safety tab.
Press the release button to make sure that the lancet is in the extended position.
Slide the ejection button forward to discard the used lancet. Place the lancing
device cover back on the lancing device.