3019 FLUORESCENCE MICROSCOPE SYSTEM
ACCU-SCOPE
®
73 Mall Drive, Commack, NY 11725 • 631-864-1000 • www.accu-scope.com
9
DETAILED ASSEMBLY
(continued)
Mercury Lamp (Fig. 5 & 6)
Loosen the lock screw
①
of the mercury lamp socket
assembly
②
on the mercury lamp housing using the M4 Allen
key, and remove the socket assembly (Fig.5).
Remove the bulb holder by loosening the lock screws
③
.
Using a soft tissue to hold the mercury lamp, insert lamp into
the (positive) “+” pole (the wide head) of the mercury burner
⑤
to the lower terminal first and then the (negative)
“–“ pole
(the thin head) to the upper terminal
⑥
, then tighten the two
socket lock screws
③
.
Slide the lamp housing cover back on, and tighten the
housing clamping screw
①
.
IMPORTANT NOTES
● Be sure to ONLY use a 100 watt mercury lamp.
● Be sure to mount the positive “+” pole (the wide head) before
the negative
“–” pole, or damage to the lamp may occur.
● Never use
excessive force when mounting the mercury lamp.
● Avoid touching or soiling the mercury lamp. If this happens,
gently wipe it with lens tissue or a soft, clean cloth.
● To prevent hazard, always turn the main switch on the power
supply unit to “O” (OFF) position, unplug the power cord plug
from the main outlet, and wait for a minimum of 10 minutes
before replacing the lamp.
Mercury Lamp House (Fig. 7)
Loosen the lock screw on the back of the fluorescence
illuminator
①
. Align and insert the connector
②
of the
mercury lamp house in the port
③
of the fluorescence
illuminator and tighten the lock screw.
The Head (Fig. 8)
Loosen the lock screw on the front of the fluorescence
illuminator
①
. Align the head with the eyepieces facing
toward the front of the illuminator and stand, and insert into
the dovetail on the top of the illuminator. Tighten the lock
screw.
Fig. 5
Fig. 6
Fig. 7
Fig. 8