3012-LED MICROSCOPE SERIES
ACCU-SCOPE
®
73 Mall Drive,
Commack, NY 11725 • 631-864-1000 • www.accu-scope.com
8
ASSEMBLY
Objectives
The objectives have been pre-installed on the microscope. Always rotate the nosepiece by using the
knurled nosepiece ring.
Viewing Head
Remove the protective plastic disk from the head and neck. Loosen the viewing head clamp screw
slightly and set the head onto the microscope arm. Tighten the thumb screw to lock the head in the
desired position. Ensure the viewing head is firmly secured by the dovetail and thumb screw before
releasing your hold.
Eyepieces
Remove the eyetube plugs and insert the eyepieces into the eyepiece tubes. Lock in the eyepieces using
the supplied Allen wrench to tighten the Allen screw on each eyetube.
Mechanical Stage
The mechanical stage is held in place by a metal retainer. Remove the two screws holding the retainer
using the supplied Allen wrench.
Condenser
The condenser was installed and centered by our technicians prior to delivery. If the condenser needs to
be re-installed or adjusted in the future, please follow the following procedure: from below, mount the
condenser on the condenser holder with the aperture control lever facing forward. Tighten the two
centering screws until the condenser is in the center of the field of view when the diaphragm is closed.
Head clamp screw for
binocular head