Meter Maintenance and Troubleshooting
7
(The error code and
Err
will alternate on the display.)
Not enough blood or control solution was drawn
into the test strip for measurement or was
applied after the test had started.
Discard the test strip and repeat the blood
glucose or control test.
(The error code and
Err
will alternate on the display.)
Blood or control solution was applied to the test
strip before the flashing drop symbol appeared
on the display.
Discard the test strip and repeat the blood
glucose or control test.
(The error code and
Err
will alternate on the display.)
An electronic error occurred.
Remove the batteries, press and hold the
meter button for at least 2 seconds, and
reinsert the batteries. Perform a blood glucose
or control test.
(The error code and
Err
will alternate on the display.)
The temperature is above or below the proper range
for the system.
Refer to the test strip package insert for system
operating conditions. Move to an area with the
appropriate conditions and repeat the blood glucose
or control test. Do not artificially heat or cool the
meter.
54461_07947836001_07947836001_01_Inside.indb 38
10/7/16 10:50 AM
Summary of Contents for WX3-122
Page 2: ...54461_07947836001_07947836001_01_COVER indb 2 10 3 16 8 02 AM ...
Page 4: ...54461_07947836001_07947836001_01_Inside indb 2 10 7 16 10 50 AM ...
Page 6: ...4 Introduction 54461_07947836001_07947836001_01_Inside indb 4 10 7 16 10 50 AM ...
Page 12: ...10 Your New System 1 54461_07947836001_07947836001_01_Inside indb 10 10 7 16 10 50 AM ...
Page 34: ...32 Connecting to a PC 6 54461_07947836001_07947836001_01_Inside indb 32 10 7 16 10 50 AM ...
Page 48: ...46 Technical Information 8 54461_07947836001_07947836001_01_Inside indb 46 10 7 16 10 51 AM ...