40
Meter Maintenance and Troubleshooting
7
Blood glucose may be higher than the measuring
range of the system.
See the
Unusual Blood Glucose Results
section
in the chapter
Blood Glucose Tests
.
Blood glucose may be lower than the measuring
range of the system.
See the
Unusual Blood Glucose Results
section
in the chapter
Blood Glucose Tests
.
(The error code and
Err
will alternate on the display.)
An electronic error has occurred.
Contact Roche.
(The error code and
Err
will alternate on the display.)
Fluid or foreign material may be present in the
test strip slot.
Remove and reinsert the test strip or repeat the
blood glucose or control test with a new test
strip. If the error persists, contact Roche.