background image

- 1 -

Operating manual

Thank you for choosing the GLUCOCARD

TM

 MX. 

We have designed this compact blood glucose meter for faster and easier 

testing of your blood glucose. We hope our product will help you manage your 

diabetes. This manual explains how to use your new meter. Before testing, 

carefully read through this manual and the package inserts that come with the 

GLUCOCARD

TM

 MX BLOOD GLUCOSE TEST STRIPS and GLUCOCARD

TM

 

MX CONTROL. Pay particular attention to listed warnings and cautions. Please 

keep this manual handy for future reference. If you have any questions, please 

contact our authorized distributor at the end of this manual.

Chapter 1  Introduction

1.1 Intended Use

Your GLUCOCARD

TM

 MX and GLUCOCARD

TM

 MX BLOOD GLUCOSE TEST 

STRIPS are for quantitatively measuring the glucose level in fresh capillary 

whole blood. They are intended for use outside of the body (in vitro diagnostic 

use) at home or in a clinical setting as an aid to monitor and control blood 

glucose levels. Do not use them to diagnose diabetes. Also, do not alter 

treatment based on the test results of this meter without instructions from your 

doctor or healthcare professional.

1.2 Measurement Principle

Glucose in the blood reacts with the reagent in the test strip and this produces 

a small electric current. The strength of this current is proportional to the 

concentration of glucose in the blood. The meter measures this current and 

calculates your glucose level. 

1.3 What’s Included

n

 GLUCOCARD

TM

 MX

n

 Carrying case

n

 Operating Manual (this manual)

n

 Simplified Manual / Information for Users

n

 Glucoject

®

 Lancets No-dol (lancets)*

n

 Glucoject

®

 Dual /Dual S (lancing device)*

Some kits do not include the above items with an asterisk (*).

Chapter 2  Before Tests

 

2.1 Important Health-Related Information 

 WARNING

Do not

 use this meter for your blood glucose testing if you are being 

tested for xylose absorption. You may use this meter if you are taking PAM 

(1-Methylpyridine-6-carbaldehyde oxime). 

These treatments may produce higher test results than your actual 

blood glucose level

. Taking unnecessary measures to lower your blood 

glucose level may induce severe low glucose symptoms such as coma. 

2.2 What You Need for Each Test

a. GLUCOCARD

TM

 

  MX (meter)

b. GLUCOCARD

TM

 MX

   BLOOD GLUCOSE 

  TEST STRIPS

c. Lancing device

d. Lancet

 WARNING

Keep the meter, test strips and other items away from young children. 

Small items are choking hazards. 

2.3 Cautions on Using the Meter

 CAUTION

n

 For accurate test results, allow the meter to adjust to the surroundings:

    Temperature: 10 to 40°C (50 to 104°F)

    Humidity: 20 to 80% RH (Relative Humidity)

  for 30 minutes before testing your blood glucose. 

n

 Do not store or use the meter where: 

  • There are sharp temperature fluctuations. 

  • Humidity is high enough to cause condensation (bathrooms, drying 

rooms, kitchen, etc.).

  • There is a strong electromagnetic field (microwave oven, cell phone, 

etc.). 

n

 Do not use the meter after it has been dropped in liquid or liquids have 

entered inside, even if dried afterwards.

n

 Keep hands away from the test strip slot on the meter. A thermo sensor 

is housed inside the meter to minimize detection of errors.

n

 Do not connect the communication cable to the connection terminal 

during testing. The meter may be damaged, leading to inaccurate test 

results.

n

 Do not apply blood directly to the test strip slot on the meter.

n

 Do not share your meter with anyone to avoid the risk of infection. 

2.4 Cautions on Using the Test Strips

 CAUTION

n

 Use only the GLUCOCARD

TM

 MX BLOOD GLUCOSE TEST STRIPS for 

testing with the GLUCOCARD

TM

 MX. 

Do not

 use other test strips as it 

causes inaccurate test results.

n

 

Do not

 use test strips beyond their expiration date. The expiration date is 

written on the test strip bottle next to “Exp.”.  

n

 For accurate test results, allow test strips to adjust to the surroundings:

    Temperature: 10 to 40°C (50 to 104°F)

    Humidity: 20 to 80% RH

  for 20 minutes before testing your blood glucose.

n

 

Do not 

use the test strips if 6 months or more have passed since 

opening the bottle. 

n

 The test strips are for single-use only. 

Do not

 use test strips that have 

already absorbed blood or control solution. 

NOTE

Store test strips in their original bottle to maintain quality. 

Do not

 transfer them 

to other containers. 

2.5 Meter Parts

Strap hole

Battery holder

Display 

with backlight

Test strip 

disposal lever

Connection

terminal

Test strip slot with light

Insert a test strip here before 

testing.

Operating 

button

MASTER

- 1 -

Summary of Contents for GLUCOCARD MX

Page 1: ...induce severe low glucose symptoms such as coma 2 2 What You Need for Each Test a GLUCOCARDTM MX meter b GLUCOCARDTM MX BLOOD GLUCOSE TEST STRIPS c Lancing device d Lancet WARNING Keep the meter test...

Page 2: ...test strip into the meter Double beeps begin to sound 4 minutes and 30 seconds after inserting the test strip Thirty seconds later the meter turns itself off NOTE In the ECO mode the blood symbol flas...

Page 3: ...d The meter counts down from 5 to 1 during the test NOTE n For accurate test results touch the test strip to your blood in less than 20 seconds after puncturing n Do not test blood that runs or spread...

Page 4: ...yond the 500th overwrite previously stored results in the order of oldest first 1 Make sure the meter is off 2 Turn on the meter The most recent test result appears on the screen The 3 digit number fo...

Page 5: ...ucose When the clock counter setting is on the elapsed time and after meal flag appear on the right of the test result 2 To change or set the elapsed time Select the time Confirm your setting 3 Slide...

Page 6: ...old To make another alarm setting repeat steps 1 to 4 To the top of the beeper volume setting To the top of the date setting The meter turns off NOTE The meter turns itself off if left alone for 3 min...

Page 7: ...toring Your Meter After use tightly close the caps of the test strip bottle and control solution bottle to maintain their quality Store your meter test strips control solutions and manuals in your car...

Page 8: ...accuracy Within 75 seconds month Operating Temperature 10 to 40 C 50 to 104 F environment Humidity 20 to 80 RH No condensation Expected life 3 years according to company data Dimensions L102 W33 18 5...

Page 9: ...Operating Manual MASTER 43125 12 11 420 x 300 mm BLACK K 100...