Remove the drum from the meter and return it to
its vial. Insert the second drum from the package
and use all 17 test strips. When it is time to
replace the drum, insert the first drum with the
damaged bar code. When the error message is
displayed, press the SET button. The meter uses
the code from the previous drum. “CTRL” flashes
in the display, requesting you to perform a quality
control check. You can use all 17 test strips. The
drum symbol continues to flash in the display until
a new drum is inserted.
All test strip drums in the same package
have a bar code unique to that package.
Do not try to code the meter using drums
from a different package. Check if you are
using the correct drums by comparing their
batch codes. The batch code is preceded
by the symbol
. If the message is still
displayed, please contact our Accu-Chek
Blood Glucose Meter Careline (see Chapter
12 “Customer support and service”).
LOT
Action
82
11
Display messages
Drum symbol and “E-2”
flash when the
first
drum
from a new pack
is inserted. During
subsequent tests the
drum symbol flashes.
Display
Cause
The bar code on the drum
is damaged.
Warning!
83