• Move the meter somewhere warmer and repeat
the test.
• Contact our
Accu-Chek Customer Careline
(see Chapter 12 “Customer support and service”).
Continue testing until all the test strips in the test
strip drum have been used. Then insert two new
batteries (type AAA) and a new test strip drum.
Insert two new batteries (type AAA).
Always store the meter at above minus 10°C.
Action
77
76
11
Display messages
Meter will not turn on.
Battery symbol appears
and disappears after
3 seconds.
Clock has stopped.
Clock is slow after the
meter is turned on.
Display
When symbols and numbers are shown in purple
below, it indicates that they flash in the meter`s
display.
Cause
• Batteries are empty.
• Ambient temperature is
too low.
• Display is damaged.
• Meter is defective.
Battery power is sufficient
for a maximum of 50
more tests.
Insufficient battery power —
no further tests can
be performed.
The meter was exposed to
temperatures below
minus 10°C, and the batteries
are starting to freeze.