18
Control Tests
2
Understanding Out‑of‑Range Control Results
w
WARNING
The control ranges are printed on the test strip container label. If the control result is out of range, check this
list to help solve the problem.
Troubleshooting Checks
Action
Were the test strips or control solutions expired?
Discard the test strips or control solution if either is
past the use by date. If the control solution was
opened more than 3 months ago, discard it. Repeat
the control test with an unexpired test strip and an
unexpired control solution.
Did you wipe the tip of the control solution bottle
before use?
Wipe the tip of the bottle with a tissue. Repeat the
control test with a new test strip and a fresh drop of
control solution.
Were the caps on the test strip container and the
control solution bottle always closed tightly?
Replace the test strips or control solution if you think
either was uncapped for some time. Repeat the
control test.
Was the test strip used immediately after it was
removed from the test strip container?
Repeat the control test with a new test strip and a
fresh drop of control solution.
Were the test strips and control solutions stored
in a cool, dry place?
Repeat the control test with a properly stored test
strip or control solution.
Did you follow the directions?
Read the chapter
Control Tests
and repeat the
control test.
Did you choose the correct control solution level,
either 1 or 2, when you performed the control
test?
If you chose the wrong control solution level, you can
still compare the control result to the range printed
on the test strip container.
Are you still unsure of the problem?
Contact Accu‑Chek Customer Care at
1‑800‑363‑7949.
52107_07111223001_EN.indb 18
4/2/15 9:22 AM
Summary of Contents for AVIVA COMBO
Page 2: ...52107_07111223001_FCBC indb 2 4 2 15 11 04 AM ...
Page 4: ...52107_07111223001_EN indb 2 4 2 15 9 22 AM ...
Page 42: ...40 Meter Settings 4 52107_07111223001_EN indb 40 4 2 15 9 22 AM ...
Page 48: ...46 Review Your Data 5 52107_07111223001_EN indb 46 4 2 15 9 22 AM ...
Page 62: ...60 Wireless Communication and Meter Pairing 6 52107_07111223001_EN indb 60 4 2 15 9 22 AM ...
Page 70: ...68 Data Transfer and PC Reports 7 52107_07111223001_EN indb 68 4 2 15 9 22 AM ...
Page 94: ...92 Technical Information 9 52107_07111223001_EN indb 92 4 2 15 9 22 AM ...
Page 96: ...94 Guarantee 10 52107_07111223001_EN indb 94 4 2 15 9 22 AM ...
Page 98: ...96 Notes 52107_07111223001_EN indb 96 4 2 15 9 22 AM ...
Page 100: ...52107_07111223001_FR indb 2 4 2 15 9 21 AM ...
Page 138: ...40 Réglages du glycomètre 4 52107_07111223001_FR indb 40 4 2 15 9 21 AM ...
Page 144: ...46 Consultation des données 5 52107_07111223001_FR indb 46 4 2 15 9 21 AM ...
Page 166: ...68 Transfert de données et rapports PC 7 52107_07111223001_FR indb 68 4 2 15 9 21 AM ...
Page 190: ...92 Informations techniques 9 52107_07111223001_FR indb 92 4 2 15 9 21 AM ...
Page 192: ...94 Garantie 10 52107_07111223001_FR indb 94 4 2 15 9 21 AM ...
Page 194: ...96 Remarques 52107_07111223001_FR indb 96 4 2 15 9 21 AM ...