When to Perform a Control Test
Performing a control test lets you know the meter and test strips are working properly. You should perform a
control test when:
• you open a new test strip box.
• you left the test strip container open.
• you think the test strips are damaged.
• you want to check the meter and test strips.
• the test strips were stored in extreme temperatures, humidity, or both.
• you dropped the meter.
• your test result does not match how you feel.
• you want to check if you are performing the test correctly.
About the Control Solutions
• Use only Accu‑Chek Aviva control solutions.
• Close the control solution bottle tightly after use.
• Write the date you open the control solution bottle on the bottle label. The control solution must be
discarded 3 months from the date the control solution bottle was opened (discard date) or on the use by
date on the bottle label, whichever comes first.
• Do not use control solution that is past the use by or discard date.
• Refer to the control solution package insert for control solution storage conditions.
• The meter automatically recognizes the difference between the control solution and blood.
• The control results are not displayed in memory.
• The control solution can stain fabric. Remove stains by washing with soap and water.
Chapter 2: Control Tests
52107_07111223001_EN.indb 13
4/2/15 9:22 AM
Summary of Contents for AVIVA COMBO
Page 2: ...52107_07111223001_FCBC indb 2 4 2 15 11 04 AM ...
Page 4: ...52107_07111223001_EN indb 2 4 2 15 9 22 AM ...
Page 42: ...40 Meter Settings 4 52107_07111223001_EN indb 40 4 2 15 9 22 AM ...
Page 48: ...46 Review Your Data 5 52107_07111223001_EN indb 46 4 2 15 9 22 AM ...
Page 62: ...60 Wireless Communication and Meter Pairing 6 52107_07111223001_EN indb 60 4 2 15 9 22 AM ...
Page 70: ...68 Data Transfer and PC Reports 7 52107_07111223001_EN indb 68 4 2 15 9 22 AM ...
Page 94: ...92 Technical Information 9 52107_07111223001_EN indb 92 4 2 15 9 22 AM ...
Page 96: ...94 Guarantee 10 52107_07111223001_EN indb 94 4 2 15 9 22 AM ...
Page 98: ...96 Notes 52107_07111223001_EN indb 96 4 2 15 9 22 AM ...
Page 100: ...52107_07111223001_FR indb 2 4 2 15 9 21 AM ...
Page 138: ...40 Réglages du glycomètre 4 52107_07111223001_FR indb 40 4 2 15 9 21 AM ...
Page 144: ...46 Consultation des données 5 52107_07111223001_FR indb 46 4 2 15 9 21 AM ...
Page 166: ...68 Transfert de données et rapports PC 7 52107_07111223001_FR indb 68 4 2 15 9 21 AM ...
Page 190: ...92 Informations techniques 9 52107_07111223001_FR indb 92 4 2 15 9 21 AM ...
Page 192: ...94 Garantie 10 52107_07111223001_FR indb 94 4 2 15 9 21 AM ...
Page 194: ...96 Remarques 52107_07111223001_FR indb 96 4 2 15 9 21 AM ...