Notes:
• Make sure that no liquid enters the meter. Do not spray anything onto the meter and do
not immerse it in liquid. Doing so may damage the meter’s internal components and stop
it from working properly.
When to Disinfect
Each time after you have tested a patient’s blood glucose: wipe the area around the test strip
slot. If you find blood on the cloth, also wipe the test strip slot itself and the display.
If there is blood on the meter: wipe the meter.
If blood has entered the meter: wipe the area around the test strip slot and the display screen.
Allow areas you have wiped time to dry thoroughly.
The meter, including the test strip slot and display screen, must be thoroughly cleaned and
disinfected at least once a day after use. Carefully check all recesses, grooves and gaps.
Suitable Disinfecting/Cleaning Solutions
Products suitable for use:
• Super Sani-Cloth
®
• 70% isopropyl alcohol
• Mild washing up liquid mixed with water
• 10% household bleach solution (1 part bleach plus 9 parts water) made the same day
63