22
19. Cleaning & Disinfection ------------------
Cleaning Information:
Cleaning must be carried out at regular
intervals as determined by the facility. The
bed must be cleaned between patients.
It is expected that cleaning the bed as
described will take 15-30 minutes.
To disinfect the bed, only use detergents
designed for use in healthcare. Do not use
abrasives, scourers or other materials that
could damage the coating. Do not use
corrosives, caustics or strong acids.
Do not use detergents that could alter the
structure or behaviour of the plastics (petrol
etc.).
Clean by wiping with a damp cloth.
The bed is not designed for maintenance in
automatic bed washers or for cleaning with
pressurised water, spraying, showering or
steam cleaning.
Accora cannot be liable for any damage or
risk of damage if inappropriate cleaning or
disinfectant agents are used.
Cleaning Procedure:
1. Remove all accessories, mattress etc.
2. Adjust the mattress platform to the
highest position and adjust the position of
the backrest and legrest to provide
access for cleaning all the platform parts.
3. Disconnect the bed from the power
supply.
4. Move the bed to where cleaning will take
place and lock the bed castors.
5. Clean
as described in the “Cleaning
Information” section above.
WARNING
The bed must be disconnected from the
power supply when being cleaned or
disinfected.
All functions MUST be tested and
approved by a competent person after
cleaning or disinfection.
The bed MUST be cleaned and
disinfected before re-using the bed for a
different patient.