13 

6.  Fit the assembled middle bracket to the 

side panel. There are two sets of fixing 
holes on each side panel, depending 
which side of the bed the side panel is 
fitted, as shown in Figure 9.4. Fit the 
bracket to the set of holes closest to the 
headboard using the M6 x 20 bolts and 
M6 washers as shown in Figure 9.5. 
Tighten the bolts using the 4mm Allen 
key.  

 

Figure 9.4 

 

Figure 9.5 

7.  Mount the side panels onto the bed frame 

as shown in Figure 9.6. Position the side 
panel between the headboard and 
footboard and fit the centre bracket claw 
over the bed frame side bar. 

 

Figure 9.6 

 

8.  Fit each side panel to the side panel end 

brackets with the M6 x 20mm bolts and 
M6 washers as shown in Figure 9.7. 
Leave the bolts slightly loose. 

 

Figure 9.7 

9.  Adjust the side panels so they are fitted 

snugly against the headboard and 
footboard.  Tighten the bolts previously 
left loose using the appropriate Allen key. 

10. Carry out the functionality check in 

Section 13. 

 

Headboard

 

Summary of Contents for EMPRESA NSB-0-FL6-000

Page 1: ...Empresa Instruction Manual IFU FL6 001EN Rev 02...

Page 2: ...anual All actions and handling of the bed must be performed in accordance with the instructions in this manual Please ensure that the manual is available to users and operators throughout the bed s se...

Page 3: ...er cable 8 Mattress guides 9 UDI Label 10 Manufacturers nameplate inc serial number 11 Legrest actuator 12 Main lift actuator Head end 13 Backrest actuator 14 Main lift actuator Foot end 15 Control Bo...

Page 4: ...us electric hazards The bed must not be used if there is any visible damage to this cable 12 12 The hand control should be positioned to avoid strangulation risk Inappropriate use of the hand control...

Page 5: ...exceeded 7 If there is a patient on the bed the bed castors must be locked as an unlocked bed castor can result in movement of the bed and injury to a patient who leaves the bed or changes position 8...

Page 6: ...disability or disease 3 Application Environment 5 Outpatient ambulatory care which is provided in a hospital or other medical facility under medical supervision where ME equipment is provided for the...

Page 7: ...perating 5 C to 40 C 41 F to 104 F 20 to 80 Non condensing 700 hPa to 1060 hPa Max 3000 meters Transport storage 10 C to 50 C 14 F to 122 F If the bed is stored in conditions outside the normal operat...

Page 8: ...ion bar WALLB 0 FL6 000 Note the bed lever cannot be fitted to the bed at the same time as the fabric siderail or the folding siderail 7 Electrical Specification Duty Cycle Intermittent operation 2 mi...

Page 9: ...ither side of the bed as shown by 1 right side and 2 left side on Figure 8 2 The bed will be delivered with the handset cable plugged into the socket in position 2 on the left side Figure 8 2 WARNING...

Page 10: ...bolts 5 24 x M8 washers 6 8 x M8 locking nuts 7 1 x 5mm Allen key 9 Position the brackets on the footboard as Figure 8 4 note the orientation of the brackets Figure 8 4 10 Fix the brackets to the foot...

Page 11: ...e locations shown in Figure 8 8 They may be fitted in any position within the locations shown Figure 8 8 Figure 8 9 17 Carry out the Functionality Check in Section 13 18 The Empresa is now ready for u...

Page 12: ...l 3 Bed 90cm wide with folding siderail fitted 4 Bed 90cm wide without folding siderail Figure 9 2 4 Fit the claw to the plate in the correct position for the bed configuration using two M6 x 20 bolts...

Page 13: ...7 Mount the side panels onto the bed frame as shown in Figure 9 6 Position the side panel between the headboard and footboard and fit the centre bracket claw over the bed frame side bar Figure 9 6 8...

Page 14: ...e sure the castor brakes are in the locked position before using the handset to change the positions of the bed Always engage the brakes when the bed is stationery or left unattended Check for obstruc...

Page 15: ...1 and button 12 together for 3 seconds the handset will beep quickly to indicate that the handset is unlocked Reset to flat and level function The bed can be returned to a position where the mattress...

Page 16: ...er between the safety stop position and the floor level position NOTE If either of the buttons are released the bed will stop moving immediately 5 When the bed has reached the floor level position sto...

Page 17: ...1 2 Slide the head foot board outwards as Figure 12 2 until the holes in the headboard bracket marked Extended in Figure 12 3 line up with the mounting holes on the bed frame Figure 12 2 Figure 12 3...

Page 18: ...s There are two types of castors fitted to the Empresa depending on model Care must be taken to make sure the castor brakes are always locked when the bed is in use being assembled or dismantled so th...

Page 19: ...use for more information Mattress size Width adjustment kit WDEX 0 FL6 000 Length setting See section 12 200cm x 90cm Standard length standard width Not required Standard setting at head and foot end...

Page 20: ...ly and all cables to prevent damage 4 Unlock the castors See Section 14 and move the bed as required 5 When the bed has been moved or repositioned lock all the castors 6 Reconnect the power supply cab...

Page 21: ...ck to enable the cable to move with the frame as it raises and lowers See Figure 18 4 Figure 18 4 Figure 18 5 WARNING When fitting accessories with a power cable to the Empresa the bed must be disconn...

Page 22: ...with pressurised water spraying showering or steam cleaning Accora cannot be liable for any damage or risk of damage if inappropriate cleaning or disinfectant agents are used Cleaning Procedure 1 Rem...

Page 23: ...ins power supply and the socket is switched on Check mains power supply connection to control box The bed is not functioning The control box indicator light is green but does not change to orange when...

Page 24: ...ns of abuse or excessive wear 3 Are all fixtures and fittings tight and secure 4 Does the bed frame appear stable and secure 5 Are all accessories fitted in line with the accessory manufacturer or acc...

Page 25: ...ust be carried out by suitably trained and qualified personnel 24 Guarantee The Empresa has a warranty period of 2 years on the electrical components and accessories and 10 years on the frame 25 Dispo...

Page 26: ...SPR 11 Group 1 The bed uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment RF emissions CI...

Page 27: ...1 0 2kV line s to earth 1kV differential mode Not applicable Mains power quality should be that of a typical commercial or hospital environment Voltage Dips short interruptions and voltage variations...

Page 28: ...ere P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer and d is the recommended separation distance in metres m Field strengths from fixed RF...

Page 29: ...according to frequency of transmitter m 150 kHz to 80 MHz d 1 2 P 80 MHz to 800 MHz d 1 2 P 800 MHz to 2 5 GHz d 2 3 P 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23...

Page 30: ...odel number Serial number Maximum patient weight Physical description of an adult Manufactured date Manufacturer Medical Device in accordance with EU Medical Device Regulation 2017 745 Floor Level fun...

Page 31: ...t dispose of in household waste Degree of protection against electric shock Class II Double Insulated For indoor use only IPX6 Degree of protection against liquid ingress EC Representative 28 Contact...

Page 32: ...32 Accora Barrington Road Orwell Cambridge SG8 5QP United Kingdom T 44 0 1223 206100 info accora care www accora care...

Reviews: