INSPECTION AND CLEANING
Perform the following inspection and cleaning procedures periodically as a
preventive maintenance measure to keep the Acclaim System in optimal
condition resulting in trouble-free operation producing crisp, clear images.
M A I N T E N A N C E
CAUTION:
Unplug the Acclaim System components before performing cleaning.
Cleaning -
Clean each component of the Acclaim System as necessary by
performing the following procedures.
1. Camera Handpiece - Following every patient use, remove the infection
control sheath from the Camera Handpiece and dispose in accordance
with local regulations. The Camera Handpiece can be wiped clean and
disinfected with any non-corrosive surface disinfectant such as Birex® se.
2. Camera Window Lens - Remove debris or spots from lens window by
cleaning with alcohol and cotton swabs.
Inspection -
Routinely inspect each component of the Acclaim System for pos-
sible defects as follows:
1. Camera Handpiece -
a.
Check overall handpiece for chips, cracks or other irregularities.
b.
Check the lens window for debris or spots.
c.
Check the connector socket for damage.
2. Cables and Connectors -
a.
Check cables for damaged or deteriorated insulation kinking or
twisting.
b.
Check connectors for loose, bent or missing pins.
c.
Check that the quick-disconnect plug end snaps close to snugly
secure the Camera Handpiece connector.
Do not spray cleaning liquids or disinfectants directly on the camera hand-
piece. Use care not to allow liquids to run into internal circuitry.
Do not wipe the surfaces using benzine, thinner, etc. as this may degrade
the finish.
Do not autoclave the Camera Handpiece.
Do not attempt any internal service of Acclaim System components. Contact your
local authorized air techniques dealer for service. Failure to heed this warning may
result in equipment damage and voiding the warranty.
2
The infection control sheath must be used for only one
patient and disposed of properly in accordance with
local code.