- 42 -
OMNISTIM
®
FX
2
PRO USER MANUAL
COPYRIGHT © 2004 - 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED
Figure 5a
Figure 5b
4.
Verify the CHARGE light is on or blinking (Figure 6).
5.
Power on the stimulator and verify that you hear a click of the safety circuit relay energizing.
Figure 6
LEAD WIRE TESTER OPERATION
Lead Wires should be inspected periodically, as well as any time there are output anomalies reported by the patient.
Plug the distal (patient end) of the lead wire into the test connectors while observing polarity (red / black).
Plug the device end of the lead wire into the receptacle labeled LEAD WIRE TESTER.
When the POWER is on, green LED will turn on if the lead wire passes the test. LED will be red if the
lead wires are faulty and need to be replaced.
When lead wires pass the test, check for loose connections by holding the lead wire about 3 inches from
plug-in connectors and slowly rotate the lead wire in a circular motion while the wires are plugged in. if
the indicator LED is flickering between red to green, the lead wire should be replaced.
Press power
supply switch
ON
For reference only:
Charge light solid - charging in
progress.
Charge light blinks slowly - charging
completed.
Charge light flickers (blinks rapidly) -
batteries not installed or are not
charging properly.
Charge light