- 34 -
OMNISTIM
®
FX
2
PRO USER MANUAL
COPYRIGHT © 2004 - 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED
E.
FAST WALK
Program simulates fast walking speed at a rate of 88 strides/minute. The program is used to re-educate gait
function. Electrodes are applied to both quads while sitting with hips and knees flexed. A light resistance
using weight cuffs or exercise band can be applied. The stimulation should cause a light to moderate
contraction. Transition to standing, with weight shifting activities then to walking in place in later rehab.
User
Adjustable
Parameter
(Y/N)
Order in
Parameter Set
Loop
Parameter
Name
Adjustable
Range (unit)
Default (unit)
Y
2
PHASE DURATION
40…100 (μs)
70 (μs)
N
CYCLE TIME
0.68 (s)
N
PULSE RATE
50 (Hz)
Y
1
TREATMENT TIME
0…99 (min)
15 (min)
Functional Re-Ed Assisted
A.
MFAC HAND CONTROL
Program provides for active assisted treatment to develop functional movement of superficial and deep muscle.
Activation of muscle with electrical stimulation occurs during functional movement. The patient should
participate to the extent possible with the functional movement by contracting during the stimulation “ON” time.
Target tissue is superficial and deep muscle groups. Bipolar set-up over agonist and antagonist muscles for target
muscle. Set intensity to elicit a grade 3-5 muscle contraction. Reduce treatment time based on muscle fatigue.
User
Adjustable
Parameter
(Y/N)
Order in
Parameter
Set Loop
Parameter
Name
Adjustable
Range (unit)
Default (unit)
Y
3
CARRIER FREQ
2.0, 2.5, 4.0, 5.0,
10.0 (kHz)
2.5 (kHz)
Y
2
BURST FREQUENCY
0.0…250 (Hz)
50 (Hz)
N
BURST RATE SCAN
20 (%)
N
RATE SCAN
ON
N
RATE SCAN TIME
4 (s)
Y
1
TREATMENT TIME
0…99 (min)
15 (min)
N
CHANNEL MODE
A/B SIM.