AH-TX Series Handheld Pulse Oximeter Instruction Manual
51
Guidance and Declaration – electromagnetic Immunity
The device is suitable for use in the electromagnetic environment specified
below. The customer or the user of device should make certain that they
are in such an environment when operating it.
Immunity test
IEC 60601 test level
Compliance level
Conducted RF
IEC 61000-4-6
N/A
N/A
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to 2.5 GHz
3 V/m
Electromagnetic environment - guidance
IEC 61000-4-11
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic
The power frequency of
magnetic fields should be at
levels characteristic of a
typical location in a typical
commercial or hospital
environment.
Note: U
T
is the AC mains voltage prior to application of the test level.